Tension Pneumothorax Decompression: 

1.

 

Locate black blood pressure bulb on right axillary side 
of the torso.  

2.

 

Tighten air release knob so that no air escapes as you 
pump bulb. 

3.

 

Pump bulb several times to prepare each of four (4) 
sites for decompression.  

4.

 

Insert either a 16 or 18 gauge IV catheter.  

 
Puncturing correct intercostal space will allow “hissing of air” 
as chest is decompressed.  

 

To repair external puncture sites: 

1.

 

Use a tongue depressor or small spatula to apply a 
thin layer of putty wax to puncture site.  

2.

 

Smooth out to desired appearance.  

 

To replace internal bladder bags:

  

 

(A)  Bilateral Mid-Clavicular Sites: 

 

Remove chest skin from torso. 

 

Lift chest plate exposing underside. 

 

Remove screws that secure Velcro

®

 strap. (Figure 1)  

 

 

Fig. 1 

 

Remove foam pads. 

 

Disconnect bladder hose from inline hose connector, 
being sure hose does not fall through hole and into 
torso. 

 

Remove bladder from site cavity and slip hose out of 
cable tie. 

260-00001 

                                           Laerdal  

7

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Laerdal Medical Documentation – Check Revision Before Use 

Summary of Contents for AT Kelly Torso

Page 1: ...Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Be...

Page 2: ...Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentati...

Page 3: ...erdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Chec...

Page 4: ...erdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Chec...

Page 5: ...cumentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Befor...

Page 6: ...ical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revisio...

Page 7: ...ng is recommended Straight or curved laryngoscope blades size 3 or 4 are recommended Manually Generated Carotid Pulse Locate red carotid pulse bulb on right side of manikin Squeeze bulb to generate de...

Page 8: ...s not fall through hole and into torso Remove bladder from site cavity and slip hose out of cable tie 260 00001 Laerdal 7 Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Docume...

Page 9: ...d subclavian 1 Attach IV bag and tubing to central vein tube located on right side of manikin It is in same location as blood pressure bulb and carotid pulse bulb 2 Fill IV bag with desired amount of...

Page 10: ...terials or workmanship from the purchaser and notification in writing of the defect at its option repair or replace any parts found to be defective or the entire product Warranty period does not renew...

Page 11: ...implied warranties whether of merchantability fitness or purpose or otherwise on the product its parts and accessories Cautions and Warnings This product contains Natural Rubber latex which may cause...

Page 12: ...Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Revision Before Use Laerdal Medical Documentation Check Rev...

Page 13: ...lan Titir 33 25 Section 33 Shah Alam Technology Park 40400 Shah Alam Selangor Darul Ehsan Tel 60 3 5122 7002 Fax 60 3 5122 5650 E mail customer service Laerdal com my New Zealand LAERDAL NEW ZEALAND L...

Reviews: