Coolant Maintenance
Periodically inspect the coolant for contamination. Replace if the coolant becomes
dirty/contaminated.
Replace filter:
As filter ages, it will become clogged with particles and contamination. Visually inspect and replace
filter as necessary. Refer to spare parts for filter part number.
Cleaning the Heat Exchanger
Cooling capacity is heavily reduced if the heat exchanger is contaminated. The heat exchanger must
be checked for contamination (particulates) regularly and be cleaned, if required.
The cleaning of the heat exchanger must be performed in accordance with the maintenance
intervals.
DANGER
The use of water/chemicals for cleaning the heat exchanger can cause short circuit and
damage the fan
The use of water/chemicals for cleaning the heat exchanger can damage the fan and result in a
short circuit. In this case persons are in danger to get an electric shock.
Do not clean the heat exchanger with water/chemicals.
NOTE
Damage to the fins of the heat exchanger due to improper handling of the unit.
Damaged fins of the heat exchanger lead to a reduced cooling capacity.
Take care not to damage the fins of the heat exchanger when cleaning the heat
exchanger.
Air pressure used to clean the fins must be controlled to prevent damage to the fins.
If the fins of the heat exchanger are not in a suitable condition, the unit must only be used
again once the damage has been rectified.
Requirements
Unit prepared for maintenance. Refer to
Preparing the Unit for Maintenance
Required Tools and Materials
Vacuum cleaner
Compressed air pistol/pressurized air spray
Hand brush
Brush
Procedure
1. Clean the condenser and grill with a vacuum cleaner, or compressed air.
2. Care must be taken not to damage the fins.
Summary of Contents for Nextreme Series
Page 1: ...Nextreme Recirculating Chiller Specification and User Manual Part Number 387004848 ...
Page 17: ...17 Component Locations Figure 2 Isometric Views of Unit Figure 3 Top View ...
Page 18: ...18 Figure 4 Front View Figure 5 Rear View Figure 6 Side View ...
Page 39: ...39 Figure 25 Sensor Wiring Low Voltage ...
Page 40: ...40 Plumbing Refrigeration Diagram Figure 26 Plumbing and Refrigeration Diagram ...
Page 43: ...43 Figure 31 Pump Performance of NRC5000 with Cooling Fluids Water and 40 EG W ...
Page 60: ...60 PCFLT Returns FLT fault or NO FAULTS FOUND PCFLT C Clears Fault Returns FLT Faults Cleared ...
Page 69: ...69 Declaration of Conformity ...