Lake Shore Model 321 Autotuning Temperature Controller User’s Manual
6-14
Service
6.9.2.4
Analog Output Calibration
Use the following procedure to perform the analog output calibration.
1. Attach test leads to TP3 (gnd1) and TP4 (Analog out).
2. Press and hold the Analog Out key until the Model 321 enters the Analog output calibration
mode.
3. Press the Up arrow key until Zero is selected.
4. With the DMM in DC voltage mode, adjust R22 until the voltage is 0.000 ±0.001 VDC.
5. Press the Up arrow key until Span is selected.
6. Adjust R18 until the voltage is 10.000 ±0.001 VDC.
7. Press Escape on the Model 321 to exit the Analog output calibration mode
8. Set power switch to off (
O
). Disconnect power cord and test resistor.
9. Replace top of enclosure and secure with four screws.
10. Replace back bezel and use 5/64 hex key to tighten two screws attaching bottom panel to unit.
6.10 MODEL 321-04 (THERMOCOUPLE) CALIBRATION
There are two Model 321-04 (Thermocouple) calibrations provided. If you have a Millivolt Voltage
Standard (with microvolt resolution) available, use the procedure in Paragraph 6.10.1. If no precision
voltage source is available, use the optional procedure in Paragraph 6.10.2.
6.10.1 Model 321-04 Calibration (With Millivolt Voltage Standard)
The Model 321-04 calibration with Millivolt Voltage Standard is presented in six parts. First is list of
test equipment in Paragraph 6.10.1.1. Second is the test setup in Paragraph 6.10.1.2. Third is the
input calibration in Paragraph 6.10.1.3. Fourth is the analog output calibration is provided in
Paragraph 6.10.1.4. Fifth is the thermocouple offset adjustment in Paragraph 6.10.1.5. Finally, the
internal thermocouple compensation calibration is provided in Paragraph 6.10.1.6. These procedures
will be performed with power supplied to the unit. Due to the presence of lethal currents, only
personnel experienced in working with live circuits should perform this procedure.
6.10.1.1 Test
Equipment
This procedure requires the following test equipment:
•
Digital Multimeter (DMM) with a 4½-digit display, capable of 4-lead resistive measurements.
•
Millivolt Voltage Standard (with microvolt resolution range).
6.10.1.2 Test
Setup
Calibration test setup is as follows:
WARNING:
Lethal line voltage is present inside the box. Only qualified personnel
should attempt to perform this procedure.
1. Set power switch to off and disconnect power cord from rear of unit.
2. Use 5/64 hex key to remove four screws attaching top panel to unit.
3. Use 5/64 hex key to loosen two screws attaching bottom panel to unit.
4. Carefully remove back bezel.
5. Slide top panel back and remove from unit.