Lake Shore Model 321 Autotuning Temperature Controller User’s Manual
Service
6-15
6. Configure the Millivolt Voltage Standard as shown below:
4 (+V)
2 (V)
++
Voltage
Standard
6.10.1.3 Input
Calibration
Use the following procedure to perform the input calibration.
1. Power on (
l
) instrument.
2. Allow instrument to warm up for at least one hour.
3. Set Voltage Standard to zero volts.
4. Attach test leads to TP8 (GND1) and TP9 (Vref adj.).
5. Adjust R52 until DMM voltage reads 2.5000 ±0.0001 VDC.
6. Configure the Model 321 to display units in millivolts.
7. Turn off temperature compensation.
8. Adjust R27 until Model 321 display reads exactly 0.000 mV DC.
9. Change Voltage Standard to +44.000 mV.
10. Adjust R43 until Model 321 display reads exactly 44.000 mV DC.
11. Switch Voltage Standard to -44.000 mV DC. The Model 321 display should read -44.000
±0.002 mV DC. If not, set Voltage Standard to zero volts and repeat steps 8 through 11.
6.10.1.4 Analog Output Calibration
Use the following procedure to perform the analog output calibration.
1. Attach test leads to TP3 (gnd1) and TP4 (Analog out).
2. Press and hold the Analog Out key until the Model 321 enters the Analog output calibration
mode.
3. Press the Up arrow key until Zero is selected.
4. With the DMM in DC voltage mode, adjust R22 until the voltage is 0.000 ±0.001 VDC.
5. Press the Up arrow key until Span is selected.
6. Adjust R18 until the voltage is 10.000 ±0.001 VDC.
7. Press Escape on the Model 321 to exit the Analog output calibration mode.
8. Set power switch to off (
O
). Disconnect power cord and test resistor.
9. Replace top of enclosure and secure with four screws.
10. Replace back bezel and use 5/64 hex key to tighten two screws attaching bottom panel to unit.
6.10.1.5 Thermocouple
Offset
Adjustment
When a new or different thermocouple is attached to the instrument, and external thermocouple
compensation is being used, it may be necessary to adjust the offset to compensate for
discrepancies in thermocouple material, leads and connections. Use the following procedure to
perform the thermocouple offset adjustment.
1. Open the Model 321 enclosure. Refer to Paragraph 6.10.1.2, Steps 1 – 5.
2. Place the thermocouple in a reference bath of known temperature (liquid nitrogen, ice, etc.) and
install external compensation. Allow the system to stabilize to the reference temperature.
3. On the front panel of the Model 321, select the desired temperature units.
4. Turn off thermocouple compensation.
5. Adjust R27 until the display reads the reference temperature.
6. Close the Model 321 enclosure. Refer to Paragraph 6.10.1.4, Steps 8 – 10.