Lake Shore Model 642 Electromagnet Power Supply User’s Manual
Computer Interface Operation
5-37
STOP
Stop Output Current Ramp Command
Input: STOP
[term]
Remarks:
This command will stop the output current ramp within two seconds of sending the command.
To restart the ramp, use the
SETI
command to set a new output current setpoint.
XPGM
External Program Mode Command
Input: XPGM
<mode>[term]
Format:
n
<mode>
0 = Internal, 1 = External, 2 = Sum.
Example:
XPGM 1[term] –
Places the Model 642 into external program mode where the output current is set by
an external voltage.
XPGM?
External Program Mode Query
Input: XPGM?[term]
Returned:
<mode>[term]
Format:
n