background image

 

Calibration procedure of O2-METER and CO2-METER 

 

www.lambda-instruments.com

 

www.bioreactors.eu

 

www.fermentor.net

 

 

Calibration procedure for LAMBDA OXYMETER (O2-Meter) 

 

The O2 concentration can be calibrated on the LAMBDA OXYMETER (O2-Meter). The 
calibration is a two-point calibration consisting of zero-setting and sensitivity setting: 

 

Zero setting 

Connect a source of nitrogen (N2) gas to the OXYMETER input barb. The gas flow rate 

should be around 200 ml/min. Unplug the power supply and continuously press the 

 

button located below the lowest order digit

 of the display and reconnect the power 

supply. The display shows 

“NUL”

 and a value, which is gradually stabilizing and 

approaching zero. When the reading is stable, press button 

P [mbar]

. The value of the 

display is saved, the display shows 

“0.00”

 and the OXYMETER is switched to operation 

mode. The saved value is subtracted from the value measured by the sensor, thus 
correcting the zero setting. This correction applies to both parameters, the concentration 
O2 [%] and the partial pressure O2 [mbar]. 

If you do not want to save this new zero setting, disconnect the power supply and 
reconnect it again (without pressing any button). The zero value is now corrected by the 
setting that was previously stored in the memory. 

If you want to disable the zero setting, unplug the power supply and continuously press the 

 button located below the lowest order digit of the display and reconnect the power 

supply. End the calibration mode by pressing simultaneously the buttons 

[ºC]

 and 

O2

 

[mbar]

. By doing this the value zero is saved in the memory and the actual sensor 

value is displayed in operation mode. 

 

Sensitivity setting

 

Connect a source of air to the OXYMETER input barb. The gas flow rate should be around 
200 ml/min. Press the 

O2 [%]

 button. The display will stabilize with the actual value of 

oxygen concentration. After the actual value of oxygen concentration has stabilized the 
correction factor can be calculated as the ratio between the real oxygen concentration 
(e.g. 21.0 %) and the actual measured oxygen concentration. 

Unplug the power supply and continuously press the 

 button located below the highest 

order digit

 of the display and reconnect the power supply. The display shows 

“SLP”

 and 

then the value of the correction factor. Use the 

 

buttons under the display to set the 

desired value of the correction factor and press button 

P [mbar]

. The correction factor is 

saved, the display shows the actual O2 concentration and the OXYMETER is switched to 
operation mode. 

Reviews: