background image

 

Calibration procedure of O2-METER and CO2-METER 

 

www.lambda-instruments.com

 

www.bioreactors.eu

 

www.fermentor.net

 

 

Calibration procedure for LAMBDA CARBOMETER (CO2-Meter) 

 

The CO2 concentration can be calibrated on the LAMBDA CARBOMETER (CO2-Meter). 
The calibration is a two-point calibration consisting of zero-setting and sensitivity setting: 

 

Zero setting

 

On the back side of the LAMBDA CARBOMETER (CO2-Meter) there are two buttons 
labeled 

N2

 and 

Air

. These buttons are used for the zero setting of the instrument. 

Connect a source of nitrogen (N2) gas to the CARBOMETER input barb. The gas flow rate 
should be around 200 ml/min. The display shows a value, which is gradually stabilizing 
and approaching zero. After the value on the display stabilizes, press the button in the hole 
on the rear side labeled 

N2

. The display shows 

“0.00”

A simpler but less accurate way is to use an air flow for the zero setting (air flow rate 
should be around 200 ml/min). In this case, after the value on the display stabilizes, press 
the button in the hole on the rear side labeled 

Air

. The display shows 

“0.04”

,

 which 

corresponds to the expected concentration of CO2 in air (400 ppm). 

 

Sensitivity setting

 

Connect a source of known concentration of CO2 to the CARBOMETER input barb. The 
gas flow rate should be around 200 ml/min. The display will stabilize with the actual value 
of CO2 concentration. After the actual value of CO2 concentration has stabilized the 
correction factor can be calculated as the ratio between the real CO2 concentration and 
the actual measured CO2 concentration. 

Unplug the power supply and continuously press the 

 button located below the highest 

order digit

 of the display and reconnect the power supply. The display shows 

“SLP”

 and 

then the value of the correction factor. Use the 

 

buttons under the display to set the 

desired value of the correction factor and press button 

. The correction factor is saved, 

the display shows the actual CO2 concentration and the CARBOMETER is switched to 
operation mode. 

Reviews: