(95666) TROPOS 

 

  

         Operating Instructions

Operating Instructions
Data logger TROPOS (95666)

Measurement and evaluation

The measurements are carried out according to the confi gura-
tion and settings (sensor signals, measuring intervals etc.), 
which are adjusted by the software 

”TROPOS-Commander

“ 

and transferred to TROPOS via RS 232.

5.1  The CF card 

Please note:

Use only CF cards which are released by LAMBRECHT -  with 
extended operating temperature range, industrial standard! 
The CF card is the central memory and should be removed 
only shortly (for max. 15 minutes) for data readout! Further-
more the operation without a CF card is an invalid operating 
mode.

A copy of the active confi guration of TROPOS is saved on the 
CF card. If no confi guration have done, the CF card contains 
the defaults station name 

TROPOS 

and station number 

0001

This actual confi guration can be transferred to the  PC via the 
button 

Request CFG 

at ”

TROPOS-Commander

.

If you want to change or adjust your confi guration of TRO-
POS by the software 

”TROPOS-Commander

“, fi rst you have 

to save this setting and afterwards you have to transfer it to 
TROPOS via the button 

Send CFG

.

Attention!

The function

 Send CFG 

may delete all data from 

the CF card!

Therefore you have to transfer all data from the CF card before 
using the function 

Send CFG

.

12) Channels and tolerances of measuring ranges

 A 

F

 A1...A2  5  k

Ω

 

±

 4 

Ω

 0.08 

OFF

 A1...A2  2.5  k

Ω

 

±

 0.5 

Ω

 0.05 

OFF

 A1...A2  250 

Ω

 

±

 0.02 

Ω

 0.02 

OFF

 A1...A6 

±

 5 V 

±

 2 mV 

0.04 % 

OFF

 A1...A6 

±

 1 V 

±

 0.5 mV 

0.05 % 

OFF

 A1...A6 

±

 100 mV 

±

 0.1 mV 

0.10 % 

OFF

 A1...A6 

±

 50 mV 

±

 0.02 mV 

0.04 % 

OFF

 A7...A8 

±

 50 mV 

±

 0.02 mV 

0.04 % 

OFF

 A1...A6 

±

 20 mA 

±

 0.02 mA 

0.10 % 

ON

 D1...D3  4  kHz 

±

 2 Hz 

0.05 % 

-

 D1...D4  status 

switching threshold

 -  - -

 

 

 

±

 50 mV

 A1...A6 

+

 10 V* 

±

 30 mV 

0.30 % 

OFF

 A1...A6 

+

 30 V** 

±

 100 mV 

0.30 % 

OFF

 

 

 

 

Table 1

* with external voltage divider... 1.5 M

Ω

** with external voltage divider... 7.5 M

Ω

A... Channel 

E... Open channel recognition

B... Measuring range 

F... Switch ON/ OFF

C... Accuracy at 23 °C
D... Accuracy in %

The memory of the CF card is suffi cient to save data 
of 1 year at least, if there are no reconfi gurations or 
exchanges of the CF card.

During CF card exchange TROPOS saves the data 
of approx. 15 minutes on the RAM. Within this time 
period the CF card should be inserted again, to avoid 
data losses.

Exchange of the CF card see 

chapter 4.3.

5.2 

 Data transfer and evaluation

The evaluation of the measured values is beginning by using 
the software 

”TROPOS-Commander

“. 

By different functions the measured values and results can 
be transferred to the PC. There the data are saved into CSV 
data fi les. By this way the data are available to processing, 
e.g. by Excel. The data transfer and evaluation can also be 
carried out by the software “

MeteoWare

” (option).

All measured values are saved according to the active con-
fi guration on the CF card and also temporary on the RAM of 
TROPOS. The data transfer is carried out by the software 

”TROPOS-Commander

“. At the menu item „Online“ differ-

ent kinds of transfer functions can be selected, by which the 
measured values can be transferred.

Further notes are available in the on-line help of the software 

TROPOS-Commander

“. The on-line help can be opened by 

using the key F1 on your keyboard for the active window.

6 Trouble-shooting

Message: TROPOS does not response.

(This message could appear only after 30...60 s!)

 

• Checking of the correct connection between TROPOS 
and PC by interface RS 232.

 

• Checking of the power supplies of both devices.

 

• Checking of the correct choice of the COM interface at 
the PC and the accordance of the baud rates.

 

• By click on the button 

ID? 

: Checking

 

of the accordance 

between the station numbers and station names from 
TROPOS with the actively chosen confi guration in the 
software 

”TROPOS-Commander

“.

The result of a data readout is a blank csv fi le

: Check the 

settings of date and time in TROPOS, concerning the readout  
time interval in

 “TROPOS-Commander”.

Dimensional and connection 

 drawings

The TROPOS ports:

See also chapter „4.4 Wiring of the TROPOS ports“, page 5 ff.

• The ports named 1...5 are located on the data processing 

module left-hand, below.

• The ports labelled 

A1...A7

 are located on the data acquisi-

tion module(s) right-hand, below. 

A8

 is located right-hand, 

above.

• The digital ports labelled 

D1...D3

, Status 

D4

 and the 4 sen-

sor power supply outputs 

OUT

 are also located on the data 

acquisition module(s) right-hand, above.

Summary of Contents for TROPOS-100

Page 1: ...very 4 Commissioning of TROPOS 4 1 Configuration of TROPOS 4 2 Operation via push rotary switch 4 3 The menu items 4 4 Wiring of the TROPOS ports 5 Measurement and evaluation 5 1 The CF card 5 2 Data...

Page 2: ...f 30 60 C The flexible structure enables a high number of measuring ranges namely voltages up to max 5 V currents up to 20 mA resistors up to 5 k as well as frequencies up to 4 kHz The inputs are prot...

Page 3: ...connecting cable null modem adapter TROPOS sided 6 Now the power supply of TROPOS may be switched on TROPOS turns on automatically and the LCD gets active The LCD turns off automatically when there a...

Page 4: ...After pressing the switch it is changing into the edit mode see above chapter 4 2 point 6 By the next rotation of the switch the menu item Date ends and the next menu item Time appears and so on Time...

Page 5: ...f TROPOS the details are described in the following text The software configuration content has to be adjusted ac cordingly The wiring drawings of the factory adjusted sensors are available after soft...

Page 6: ...und referred single ended The reference ground is the input ground on the ports J at the channels A1 and A2 as well as the ports U at the channels A3 A8 All ground ports have to be connected next to t...

Page 7: ...cies pulses or status The digital measuring inputs D1 D4 work ground referred The switching threshold is set to 2 5 V Frequencies up to 4 kHz pulses or status are measurable by the digital channels D1...

Page 8: ...card During CF card exchange TROPOS saves the data of approx 15 minutes on the RAM Within this time period the CF card should be inserted again to avoid data losses Exchange of the CF card see chapter...

Page 9: ...of them switchable via software 2 1 output Relay to switch a communication unit modem A1 A2 2 analogue inputs Resistors up to 250 or e g Pt100 in 2 or 4 wire circuit Resistors up to 2 5 k or e g Pt100...

Page 10: ...location independent data evaluation range of application 40 85 C industrial standard 1x incl in the scope of delivery of TROPOS Add on modules for data transfer TROPOS RS 485 not for solar operation...

Page 11: ...configuration If the mounting is possible the TROPOS Installation 2 is necessary 3 At external installations additional accessories or assembly works could be necessary cables wirings protective devic...

Page 12: ...min Extreme values For each average value interval wind peaks per minute Ring buffer For 1 year Memory Compact Flash card CF card Display Easy to read 2 lines 16 characters Operation Via software TROP...

Reviews: