6
TESTING
12:14 AM
30-Jan-19
Water type
Disk Series
Fresh
FF103
Tap to
Start Test
1. Press and hold until the meter turns on.
2. Tap . Select a Water type. Tap
to confi rm.
3. Tap . Select a disk series (found on disk packaging). Tap
to confi rm.
NOTE: Disk Series are limited by Water type selection.
4. Remove a disk from the packaging.
5. Use the syringe (1189) to fi ll the disk with the water sample.
6. Insert the disk. Cover the disk with the Universal Disk Cover (1719). Close the lid.
7. Tap to start test. Tap
to cancel the test. If the test is cancelled discard the
disk.
8. The results will be displayed.
9. Choose an option.
¬
Tap
to add a tag.
¬
Tap the highlighted
to save the test results to the test log if Auto Save is
not enabled.
¬
Tap the hightlighted
to send the results to the enabled Mobile Bluetooth
Printer.
¬
Tap the highlighted
to send the results to a Bluetooth enabled device.
¬
Tap
to return to the Test Screen.
10. Press and hold
for 2 seconds to turn the meter off .
For the most accurate results samples should be at room temperature.
The blank well allows the readings to be corrected for small amounts of color and turbidity in the
sample water. For the best results allow samples with a large amount of solids to settle before testing.
NOTE: For water samples over 100 °F (38 °C) subtract 0.3 from pH result or, for the most accurate
result, wait until water sample is below 90 °F (32 °C) to test.
Remove salt residue daily. Salt will damage the meter and cause inaccurate results. See Cleaning,
page 18.
GENERAL OPERA
TING PROCEDURES