SETUP & EDIT SEQUENCES &
USER TESTS
The
EDIT
menu allows any of the three alterable test sequences (
SEQUENCE 1
,
SEQUENCE 2
, and
SEQUENCE 3
) and any of the 25 User Tests in the
ALL
TESTS
fixed sequence to be edited. This feature allows a sequence or test which
is used frequently to be set-up for easy access. The order of the sequence can be
arranged to suit the needs of the user. Any combination, and any order of tests
from
ALL TESTS
(including User Tests), may be placed into these sequences.
n
EDIT A SEQUENCE
Go to the
PROGRAMMED TESTS
menu. Move the
*
down using
t
button until the
*
is to the left of
EDIT
. Press the
ENTER
/
Q
button
PROGRAMMED TESTS
SEQUENCE 3
ALL TESTS
* EDIT
The
EDIT
menu appears.
EDIT
* EDIT SEQUENCE 1
EDIT SEQUENCE 2
EDIT SEQUENCE 3
30
SMART SPECTRO SPECTROPHOTOMETER 05.04