TROUBLESHOOTING
Problem
Check
Action
Reading is frozen
HOLD mode
Press MODE/HOLD button
to exit HOLD mode
“BAT” message
Batteries low
Replace batteries
Meter will not
calibrate in
conductivity mode
Trapped air
bubbles
Tap probe or stir sample to
release air bubbles
Dirty probe
Clean conductivity probe
Damaged probe
Replace probe
Contaminated
conductivity
standards
Use fresh standards
Meter will not turn on Batteries low or
dead
Replace batteries
Battery polarity
Replace batteries with
correct polarity
Unit will not respond
to any key presses
Internal fault
Perform hard reboot.
Remove batteries, hold
ON/OFF button down for 5
seconds, replace batteries
"_oL_" message
Sample is out of
range
Dilute sample
18