Calibration Reminder

The “CAL” indicator will appear when the TRACER is in the pH mode 

and a calibration is required. The “CAL” indicator will appear if the 

meter has not been calibrated after 15 on/off cycles of the meter. Some 

applications may require more frequent calibrations than others. The 

“CAL” indicator is only a reminder and will not affect function in any 

way. The indicator will no longer be displayed when the pH electrode is 

recalibrated.

RENEW Indicator

fl

 ashing “RENEW” indicator will appear on the display to warn that 

the electrode is not performing to expected speci

fi

 cations. If cleaning 

or recalibration does not cause the RENEW indicator to disappear, the 

electrode should be replaced. The RENEW indicator will appear when 

the pH electrode slope falls below 70% of a nominal slope.

Conductivity (Salt & TDS) Calibration

Meter accuracy veri

fi

 cation should be performed on a monthly/periodic 

basis as needed. The frequency of the veri

fi

 cation will depend on the 

storage and maintenance conditions and the amount of meter use. If 

calibration is required, the meter must be in the conductivity mode to 

perform all calibrations for conductivity, TDS and salinity. The meter 

can perform a calibrations and store the data. The automatic calibration 

recognition procedure will recognize conductivity standard of 12,880

μ

S. 

(See Page 6). Always calibrate in the range closest to the expected 

measurement value. For salinity samples within the range of 1.00 to 9.99 

ppt salinity, calibrate with a 12,880

μ

S calibration standard.

1.  Fill a sample cup to 20 mL line with a conductivity standard.
2.  Press the ON/OFF button to turn the TRACER on. Insert the 

electrode into the standard. Tap or stir the sample with the Tracer to 

dislodge air bubbles.

3.  Press and hold the CAL/RECALL button for approximately 2 

seconds until the display begins to 

fl

 ash.

4.  The meter will automatically recognize and calibrate to the 

conductivity standard. The display will brie

fl

 y indicate “SA” and 

“End” and then return to the measurement mode. 
NOTE: “SA” will not appear if the calibration fails.

Summary of Contents for Tracer pH/TDS/Salt/Temp PockeTester

Page 1: ...ofessional s Meter Tracer pH TDS Salt Temp PockeTesterTM Code 1766 KIT 01 Do Not allow pH sensor to dry out Store Upright Before rst use tap meter against palm to move electrolyte to tip See pages 10...

Page 2: ...WARNING This set contains chemicals that may be harmful if misued Read cautions on individual containers carefully Not to be used by children except under adult supervision...

Page 3: ...tomatic Calibration 8 TDS Conversion Ratio 8 Changing the Display Temperature Units 9 Data Hold 9 Auto Power Off 9 Low Battery Indicator 9 Testing Getting Started 10 Changing the Measurement Function...

Page 4: ...ON Congratulations on your purchase of the pH TDS SALT TRACER PockeTester The TRACER is a revolutionary first of its kind measurement device that offers direct reading of pH TDS and salinity with one...

Page 5: ...00 to 9 99 ppt S fixed ratio of 0 5 TDS Ratio 0 4 to 1 0 adjustable Salinity Ratio 0 5 fixed Conductivity ATC 2 0 per C Conductivity ATC Range 32 0 to 140 0 F 0 0 to 60 0 C Temperature Range 23 0 F to...

Page 6: ...ets pH 7 0 Code 3984A J Conductivity Standard 12 880 S 120 mL Code 6317 J Not sold in this quantity See below PARTS ACCESSORIES pH TDS SALT Replacement Electrode Code 1755 Weighted Stand w Sample Cups...

Page 7: ...RECALL button calibration change temperature units recall data 5 ON OFF button 6 Electrode Collar 7 Electrode NOTE The Electrode cap is not shown TRACER DISPLAY 1 Bar graph display 2 Measurement units...

Page 8: ...uish TDS Conversion Ratio The TDS value is determined by multiplying the conductivity measurement by a known conversion ratio factor The meter allows the selection of a conversion ratio factor in the...

Page 9: ...he last button was pressed To disable the auto off feature 1 Press the ON OFF button to turn the meter on 2 Wait for the SELF display screen to appear This is the second screen to appear after turning...

Page 10: ...ength of time The crystals will dissolve with soaking or they can be rinsed off with tap water 5 Calibrate with pH 7 buffer prior to the first use or after prolonged storage 6 When the meter is calibr...

Page 11: ...s the ON OFF button 8888 and then SELF CAL will appear in the display during the initial diagnostics 4 Press and hold the MODE HOLD button to scroll to the desired measurement mode 5 Slowly stir the s...

Page 12: ...e from the pH reference electrode Also increase the sample size to 200 to 500 mL to decrease the chance of contamination If a sample cup is used do not allow the probe to sit in the sample for any lon...

Page 13: ...exit the HOLD function by pressing the MODE HOLD button 1 Press the CAL RECALL button and then press the MODE HOLD button The location number 1 through 25 will briefly appear and then the value store...

Page 14: ...he ON OFF button to turn the TRACER ON 3 Place the electrode in the buffer solution Press and hold the CAL RECALL button until CAL appears in the lower display The meter will automatically recognize t...

Page 15: ...eriodic basis as needed The frequency of the verification will depend on the storage and maintenance conditions and the amount of meter use If calibration is required the meter must be in the conducti...

Page 16: ...the calibration data for the currently selected range will be replaced NOTE As with all combination pH conductivity probes at low conductivity levels the flow of the pH electrolyte may affect the read...

Page 17: ...thin 4 seconds Scroll Stored Readings On Memory Recall Momentary press of MODE HOLD button Displays last in first out Exit Memory Retrieval On Memory Recall Momentary press of CAL RECALL button Clear...

Page 18: ...en momentarily press ON OFF button Release CAL RECALL button after Self Cal is displayed Override Auto Power Off On Any Measurement Mode Momentarily press CAL RECALL button then simultaneously press a...

Page 19: ...rements at low conductivity Electrode Cleaning Recommendations Do not soak the electrode in the solutions for longer than the recommended length of time To do so may cause a reference potential shift...

Page 20: ...o side while pulling it away from the meter until it disconnects from the electrode socket 3 To attach an electrode align the slots and carefully plug the electrode into the meter socket CAUTION Take...

Page 21: ...minated pH buffers Use fresh buffers Meter will not calibrate in conductivity mode Trapped air bubbles Tap probe or stir a sample to release air bubbles Dirty probe Clean conductivity probe Damaged pr...

Page 22: ...harges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of...

Page 23: ...COMPANY Helping People Solve Analytical Challenges PO Box 329 Chestertown Maryland 21620 USA 800 344 3100 410 778 3100 Outside USA Fax 410 778 6394 Visit us on the web at www lamotte com 1766 KIT MN 0...

Reviews: