ACS set „top value”
Menu selection:
Menu D1
>Disp. Device : ACS
press soft key „select“
Menu D2.1
>ACS - Settings
press soft key „select“
Display
D3.2
„ACS - Settings“ appears.
Choose value „Set Top“ using the soft keys (
45
) „up“ and „down“ .
Press soft key „select“.
By scrolling with the soft keys (
45
) „up“ and „down“ all values from the chapter „read ACS values“ appear
consecutively in the display.
Pressing the soft key „set“ selects the chosen value as „top value“.
Set up Air Flow Value
Menu selection:
Menu D1
>Disp. Device : ACS
press soft key „select“
Menu D2.1
>ACS - Settings
press soft key „select“
The air flow value must be set accordingly to the mounted flow module. Factory
setting is already appropriate.
Display
D3.2
„ACS - Settings“ appears.
Choose value „flow max“ using the soft keys (
45
) „up“ and „down“ .
Press soft key „select“.
Use softkeys (
45
) „up“ and „down“ to change between 5000 and 10000 l/h
For RTS 1000 set value to 5000 l/h.
Press softkey „set“ to store the selected value.
Service mode is closed automatically after 10 minutes.
ACS - Settings
>Set Top : Operating
flow max : 10000 l/h
Operating : 12345 h
back
______________________
up
|
down
| select
ACS - Settings
Set Top : Operating
>flow max : 10000 l/h
Operating : 12345 h
back
______________________
up
|
down
| select