background image

 

 
Ultraheat / Ultracold T550 (UC50…) 

Subject to change without prior notice 

 

Page 1 / 3 

Calculator T550 Ultraheat

®

       (UC50

…) 

Calculator T550 Ultracold

®

       (UC50

…) 

Operating Instructions 

UC 309-101d

 

 

Issue: 29.10.2012 

 

These Operating Instructions must be handed to the final user on start-up! 

Note: In the following text, the term calculator covers 
the  heat  meter  calculator  as  well  as  the  cold  meter 
calculator and the combined heat/cold meter calcula-
tor, if not mentioned otherwise.  

General information 

The  T550  Ultraheat

®

 

(UC50…)  calculator  is  de-

signed  for  heating  and  cooling  applications.  The 
calculator  forms  the  volumes  from  the  pulses  of  the 
flow  sensor.  The  flow  and  return  temperatures  are 
sensed using platinum resistors. 

The volume of water and the temperature difference 
between the flow and return are then calculated and 
the product is summated. 

The result is that the quantity of thermal energy con-
sumed  is  displayed  in  the  units  kWh / MWh or MJ / 
GJ. 

 

 

The 

mounting  place

  and  the 

pulse  value

 

of flow sensor with pulse output must corre-
spond  with  the  in  the  calculator  set  values 
(see LOOP 2).  

Technical data 

Environment class 

A (EN 1434) 

Mechanical class 

M1  *) 

Electromagnetical class 

E1  *) 

*) according 2004/22/EG directive 

Ambient humidity  

<  93 % r. h. at 25°C 

 

 

without condensation 

Storage temperature 

- 20 to 60°C 

Max. Height 

2000 m above MSL 

Ambient temperature 

5 to 55°C 

 

Housing degree of prot.  IP 54 per EN 60529 
Safety class 

 

  line 110 / 230 V AC 

II per EN 61558 

  line 24 V AC/DC 

III per EN 61558 

Response threshold f. 

T 0.2 K 

Temperat. diff. 

3 K to 120 K  

Temperat.meas. range 

0...180°C 

Sensors Type 

Pt500 or Pt100 

 

per EN 60751 

Temp. sensors length 

max. 10m 

Battery 

 

 

3,6V DC Lithium 

Power supply 

 

230/110V AC 50/60Hz 

 

   

 

 

24V AC/DC 50/60Hz 

Power input   

 

<  0,8W 

Pulse input   

 

IB/IC per EN1434 

Pulse length  

 

min. 10ms 

Pulse frequency 

 

max 50Hz 

Pulse wire length  max. 20m (recommended) 

Operating elements 

 

Displays 

The  displays  of  the  meter  are  arranged  on  several 
levels (LOOPs).  

Button 1 advances the display to the next loop. After 
the last loop, the user loop (LOOP 0) appears again.  

Button 2 displays the data of the selected loop. With-
in a loop, the button 1 is used to advance to the next 
line  of  the  display.  After  the  last  line  of  the  display, 
the first display line appears again. 

The digits after the decimal point of displayed values 
are indicated by a surrounding border.  
Calibrated  values  can  be  recognized  by  the  star 
symbol shown in addition to the value. 

 

 

 

Depending on how the unit is parameterized, 
the number of items displayed and the data 
shown may differ from this description. Cer-
tain button functions may also be disabled. 

Service loop (selection) 

 

Service loop 1 

 

Service loop 2 

 

 

 

User loop 0 

 
User loop 

(„LOOP 0“) 

 

Head of the loop 

 

Error message with error code numer 

1

)

 

 

Accumulated quantity of energy with tariff status 

 

Tariff register 1,2,3

 2

)

 

 

Accumulated volume 

 

 

 

Volume pulse input 1 

3

)

 

at 2 s interval with current volume 

 

 

 

Volume pulse input 2 

3

)

 

at 2 s interval with current volume 

 

Segment test 

 

button 1 

button 2 

LCD display 

3280 

009 

10

Reviews: