13

Limited Consumer Product Warranty 

This product is warranted to the original consumer purchaser (the “purchaser”) to be free from 
defects in material and workmanship which are not commercially acceptable for the period of 
one year from the date of purchase on the pump mechanism and the period of 90 days on 
all other product components. This warranty gives you specific legal rights and you may also 
have other rights, which vary, by location. Your statutory rights are unaffected by this warranty.

LANSINOH MAKES NO EXPRESS WARRANTY OF ANY KIND REGARDING THE PRODUCT 
OTHER THAN THOSE WARRANTIES EXPLICITLY SET FORTH HEREIN. ANY IMPLIED 
WARRANTY, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR ANY 
IMPLIED WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE, TO THE EXTENT 
PERMITTED BY LAW, SHALL BE LIMITED IN DURATION TO A PERIOD OF 90 DAYS FROM 
THE DATE OF PURCHASE BY THE ORIGINAL PURCHASER.

In the event the product is found by Lansinoh to not meet the above limited warranty, as 
purchaser’s sole and exclusive remedy Lansinoh will repair or, at the option of Lansinoh, 
replace this product without charge for such replacement or parts or labour. This warranty 
does not apply to any product that has been subjected to misuse, abuse, alteration or for 
normal wear and tear. Only Lansinoh brand replacement parts are recommended for use with 
Lansinoh breastfeeding products. Improper or incorrectly performed maintenance or repair, or 
use of non-Lansinoh brand replacement parts may limit or invalidate this warranty.

TO THE EXTENT PERMITTED BY LAW, LANSINOH SHALL NOT BE LIABLE FOR ANY 
INCIDENTAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, 
REPLACEMENT COSTS RESULTING FROM THE BREACH OF ANY WRITTEN OR 
IMPLIED WARRANTY.

If you wish to make a claim under this warranty, please telephone 0113 259 1425.  
You will be given a Freepost address to return the product, if necessary, together 
with your original dated bill of sale or other dated proof of purchase and a brief 
statement of the problem.

Warranty

Troubleshooting The Lansinoh Affinity Breast Pump
Interruption of Electrical Service

If you experience an interruption in the normal operation of the electrical and/or mechanical 
functioning of the pump, simply do the following:
• Unplug the AC adapter cord from the electrical outlet.
• Wait a few seconds, and then plug the AC adapter cord back into the electrical outlet.

Discomfort While Pumping

If you feel any discomfort, you may need a different size flange. 
Visit www.lansinoh.co.uk/customerservice or call 0113 259 1425. 

Decreased Suction

If it feels like the suction has decreased, try the following:

• 

Set cycle to its highest level (L6). The highest cycle setting creates a slower  

  pumping rhythm, which combined with the highest suction level (L8), allows the  
  Affinity breast pump to operate at its greatest level for milk expression.

•  Remove the tubing and tubing connector from the back of the pump and place a finger 
  over the pump outlet. If you feel suction, the pump is operating properly, but the 
  Breastmilk Expression Set may not be correctly assembled or the white valve may be 
  damaged. If no suction is felt on your finger, try unplugging the adapter from the electrical 
  outlet, wait a few seconds, and plug it back into the electrical outlet.
•  Make sure the Breastmilk Expression Set has been properly assembled. Check the 
  assembly information in the section “Assembly Instructions”, page 7.
•  Inspect the white valve for a tear or pinhole. This component is critical to achieving proper 
  suction. Replace the white valve with one of the spare valves provided with the product.
•  When single pumping, check the tubing adapter on the back of the pump to make sure it 
  is closed for single pumping. If open, close the adapter.
•  If double pumping, ensure that the breastmilk expression set is on both breasts before 

 

  starting the pump.

Increased suction level for single pumping:

If it feels like the suction is higher than usual when single pumping, check that the plug for the 
pipe is clean and that the airway in the plug is not blocked.

Breastmilk Build-up in Breast cushion or Breast cushion Body

If breastmilk does not appear to drain fast enough into the bottle, reduce the suction setting 
on the pump to allow the milk to drain. Once milk has drained into the bottle, you may 

gradually increase the suction level.

14

Troubleshooting

Summary of Contents for AFFINITY

Page 1: ...Bisphenol A FREE Double Electric Breast Pump Custom Expression Technology ComfortFit Breast Cushions Independently Adjustable Cycle and Suction Levels Easy to Use Easy to Clean...

Page 2: ...uipment into an outlet on a circuit different from that to which the other device s are connected Consult the manufacturer or field service technician for help When using electrical products especiall...

Page 3: ...2 3 4 5 Cycle indicator Time level indicator Low battery indicator Let down Stimulation Phase indicator Expression Phase indicator Suction indicator Tubing Adapter Input 6 7 8 9 10 11 Bottle holder 2...

Page 4: ...To First Use 1 Prior to using the set for the first time disassemble and boil all parts of the Breastmilk Expression Set for 10 minutes Note Ensure you use enough water to keep parts afloat while boi...

Page 5: ...o be helpful Some mums can achieve more than one let down during a pumping session You may be able to achieve a second or third let down during your pumping session by pressing the Quick Change button...

Page 6: ...on to increase the cycle length which creates a slower pumping rhythm for more powerful milk expression Note The highest cycle setting L6 actually creates the slowest pumping rhythm while the lowest c...

Page 7: ...der cool running water gradually warming the water until the milk is thawed Thawed milk can be kept in a refrigerator for up to 24 hours NOTE DO NOT refreeze milk once it is thawed Gently shake thawed...

Page 8: ...rruption of Electrical Service If you experience an interruption in the normal operation of the electrical and or mechanical functioning of the pump simply do the following Unplug the AC adapter cord...

Page 9: ...ass A Complies Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Electromagnetic environment guidance The...

Page 10: ...7 Protection Class II ISO 13485 ISO 9001 CE 1007 EN 980 EN1041 EN14350 for part numbers 50414 50406 50413 Operation Transport Storage 5 C 40 C 20 C 50 C Operation Transport Storage Electromagnetic Com...

Page 11: ...se II you may adjust cycle level by pressing or next to cycle icon The highest cycle setting L6 actually creates the slowest pumping rhythm while the lowest cycle setting L1 creates the fastest pumpin...

Reviews: