Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Breast Pump
This electric breast pump is intended for use in electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of the electric breast pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the electric breast pump as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150kHz to 80MHz d= 1,2
√
P
80MHz to 800MHz d= 1,2
√
P
800MHz to 2,7 GHz d= 2,3
√
P
0,01
0,06
0,04
0,07
0,1
0,18
0,11
0,22
1
0,58
0,35
0,22
10
1,84
1,11
2,21
100
5,83
3,50
7,00
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1
at 80 MHz and 800 MHz the separation distance for the higher frequency range applies
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the breast pump including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
Using other cables and accessories that not provided from Lansinoh® may negatively affect EMC performance.
12.2 FCC Statement
This device complies with part 15 of the FCC rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause undesired operation.
12.3 Wireless Statement
Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this equipment
and should be kept at least a distance of 3.5m away from the equipment.
12.4 Stacking
This device should not be used adjacent to or stacked with other equipment. And if adjacent
or stacked use is necessary, this device should be observed to verify normal operation in the
configuration in which it will be used.
12.5. Disposal
The symbol which can be found on the product or on the package surface indicates
that the product mustn’t be disposed with other household wastes.
Waste equipment disposal by handing it over predetermined collection point for recycling is
solely users’ responsibility. Seperate collection of the waste equipment and recycling them
properly will preserve the natural resources, have a positive affect on the environment and
will protect human health. For further information about the recycling collection points,
please contact your local authority or place you purchased the product.
Serious events experienced from using this product should be reported immediately to
Lansinoh Laboratories and to the local competent Authority.