Radiated Emission Test Level
Frequency range
Quasi-peak
dB(¨V/m)
30 – 230
230 – 1 000
40
47
Conducted Emission Test Level
Frequency range
MHz
Quasi-peak
dB(¨V/m)
Average
dB(¨V/m)
0,15 – 0,50
66
Decreasing linearly with logarithm
of frequency to
56
56
Decreasing linearly with logarithm
of frequency to
46
0,50 – 5
56
46
5 – 30
60
50
At the transition frequency, the more stringent limit shall apply.
Immunity Test Level for Input A.C. Power PORT
Phenomenon
Basic EMC
standard or test
method
IMMUNITY TEST LEVELS
Professional
healthcare facility
environment
HOME HEALTHCARE
ENVIRONMENT
Electrical fast
transients/bursts
a) l) o)
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
Surges
a) b) j) o)
Line-to-line
IEC 61000-4-5
± 0,5 kV, ± 1 kV
Surges
a) b) j) k) o)
Line-to-ground
IEC 61000-4-5
± 0,5 kV, ± 1 kV, ± 2 kV
Conducted
disturbances
induced by RF fields
c) d) o)
IEC 61000-4-6
3 V
m)
0, 15 MHz – 80 MHz
6 V
m)
in ISM bands
between 0,15 MHz and 80
MHz
n)
80 % AM at 1 kHz
e)
3 V
m)
0, 15 MHz – 80 MHz
6 V
m)
in ISM and amatuer
radio bands between 0,15
MHz and 80 MHz
n)
80 % AM at 1 kHz
e)
Voltage dips
f) p) r)
IEC 61000-4-11
0 % UT ; 0,5 cycle
g)
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
q)
0 % UT ; 1 cycle
and
70 % UT ; 25/30 cycles
h)
Single phase: at 0°
Voltage Interruptions
f) i) o) r)
IEC 61000-4-11
0 % UT ; 250/300 cycles
h)
Manufacturer: Lansinoh® Laboratories Saglik Gerecleri Tasarim San. Tic. Ltd. Sti.
Zafer SB Mah. Gunduz Sk. No:13 E.S.B Gaziemir/ Izmir/ Turkiye
Harmonic Emission Test Level Parameters
Table 1 – Limits for Class A Equipment
Harmonic order
n
Maximum permissible
harmonic current
A
Odd Harmonics
3
5
7
9
11
13
15
≤
n
≤
39
2,30
1,14
0,77
0,40
0,33
0,21
15
0,15—
n
Even Harmonics
2
4
6
8
≤
n
≤
40
1,08
0,43
0,30
8
0,23—
n
Flicker emission test level parameters
(Short-term Flicker)
:
■
1.00
(Relative steady-state voltage change) :
■
3.30 %
(Maximum relative voltage change)
:
■
4.00 %