Operating Instructions

 

 

Valid from: 

23.09.2020 

EPIC

®

 CRIMPING TOOL DIGITAL  

 

Document: L11148001EN 

Page 6 of 10 

We reserve all rights according to DIN ISO 16016.

 

 

7. Testing calibration to 1.0 mm or 2.0 mm using the gauge 

 

Check the default setting of your crimper (1.0 mm or 2.0 mm crimp) before use. 

 

Switch on your crimper by pressing the ON/OFF switch (7). Set to the default setting (1.0 mm or 2 mm) 
using the adjustment wheel (8). Make sure that the gauge measurement is always taken from a larger 
value, such as from 1.05 mm down to 1.0 mm or 2.05 mm down to 2.00 mm. 

 

Close the crimper and place the 1.0 mm or 2.0 mm gauge between the dies. Make sure: 

o

 

That you can move the gauge between the dies without play. If there is no deviation in 
measurement, you can use the crimper immediately. 

o

 

If you can move the gauge between the dies with play, or you cannot insert the gauge into the 
crimper, you have a deviation in measurement and will need to recalibrate the crimper.

 

8. Recalibrating the crimper (REC) 

 

Set the display to mm using the MODE button (6) (see 2. Operation). 

 

Set the crimper die using the adjustment wheel (8) until gauge supplied with the crimper touches the dies 
and you can move it without play. 

 

Make sure that the gauge measurement to be set is always taken from a larger value, such as from 1.05 mm 
down to 1.0 mm or from 2.05 mm down to 2.0 mm gauge. 

 

Keep the ON/OFF switch (7) pressed and press the MODE button (6) using the gauge. Keep the MODE 
button (6) pressed for at least ive seconds. 

 

Release the MODE button (6) after ive seconds, and then release the ON/OFF switch (7). 

 

The digital display will automatically show a gauge value of 1.0 mm or 2.0 mm. 

 

Your crimper is recalibrated and ready for crimp parameter setting. 

9. Calibrating your tool (CAL) after changing the battery 

 

Open the battery compartment (1) upwards. 

 

Remove the used battery. 

 

Turn the adjustment wheel down to the lower setting (minus sign turning direction) and leave it there. 

 

Insert the new battery. The display (5) will show CAL as a calibration request. 

 

Set the crimper to 1.0 mm or 2.0 mm using the gauge – turn the adjustment wheel (8) until you can move 
the gauge between the dies without play as shown in 7. 

 

Keep the ON/OFF button (7) pressed, and press the MODE button (6) with the gauge. 

 

Make sure that the gauge measurement is always taken from a larger value, such as from 1.05 mm down to 
1.0 mm or from 2.05 mm down to 2.0 mm. This means, that at the beginning of the adjustment the gauge 
can be inserted with having some play. 

 

Keep the MODE button pressed for at least five seconds. Release the MODE button after five seconds, and 
then release the ON/OFF switch. 

 

The digital display will automatically show a gauge value of 1.0 mm or 2.0 mm. 

 

Your crimper is calibrated and ready for crimp parameter setting. 

 

 

Summary of Contents for EPIC 11148001 EPIC 11148004

Page 1: ...ent L11148001EN Page 1 of 10 We reserve all rights according to DIN ISO 16016 EPIC CRIMPING TOOL DIGITAL EPIC CRIMPING TOOL DIGITAL SMALL Operating Instructions for 11148001 11148004 Important notes f...

Page 2: ...mping procedure 5 4 Changing the contact bushing 5 5 Changing the battery 5 6 Work process safety 5 7 Testing calibration to 1 0 mm or 2 0 mm using the gauge 6 8 Recalibrating the crimper REC 6 9 Cali...

Page 3: ...ts in plastic case without locator 0 08 2 5 28 13 175 820 This tool allows you to check it at testing intervals you set yourself and recalibrate as necessary The crimper features wear monitoring to in...

Page 4: ...8 76 3 Briefly press the MODE button 6 with the gauge provided to change the setting This will show the various display modes in the following order Setting the crimping parameters Refer to the adjus...

Page 5: ...ften you use the crimper You will need to change the battery after this period Open the battery compartment upwards in the direction of the arrow for easy battery removal and replacement Info Always s...

Page 6: ...such as from 1 05 mm down to 1 0 mm or from 2 05 mm down to 2 0 mm gauge Keep the ON OFF switch 7 pressed and press the MODE button 6 using the gauge Keep the MODE button 6 pressed for at least ive s...

Page 7: ...fe Reference value lower setting as specified by the manufacturer Number of calibrations so far 12 Troubleshooting and remedy Display Cause Solution E1 The crimper was not turned down to its lower ref...

Page 8: ...1 1 mm female hyperboloid contact 74034500 74034501 0 14 26 0 80 0 34 24 0 90 0 50 22 1 10 0 75 20 1 15 1 00 18 1 15 LS 1 2 mm male contact 74034100 74034101 0 14 26 1 25 6 0 34 24 1 30 0 50 22 1 30...

Page 9: ...t No mm2 AWG Adjustment wheel Locator D Sub 1 mm male contact 44423357 0 25 24 1 00 3 0 34 22 1 05 0 50 21 1 10 0 75 20 1 15 1 00 18 1 20 D Sub 1 mm fremale contact 44423356 0 25 24 1 00 4 0 34 22 1 0...

Page 10: ...11148001 Operating Instructions Valid from 23 09 2020 EPIC CRIMPING TOOL DIGITAL Document L11148001EN Page 10 of 10 We reserve all rights according to DIN ISO 16016 M17 contacts 11148004 11148304...

Reviews: