8-20
Model 820 Technical Reference Manual
152
E. A. Cal Tone
[No Yes]
Automatic electrostatic actuator cali-
bration tone control (820 OPT-01
must be installed).
153
Heater On
[Yes, No]
154
Modem Mode
[No, Yes]
Select if Model 820’s RS-232 port is
connected to modem
155
Deal Out Mode
[None, EXCD, ALRM,
Both]
Instrument automatically dials out if:
EXCD: RMS Exceedance level 2 is
Exceeded
ALRM: Enclosure intruder alarm is
triggered or
Both.
156
Phone #
157
Monitor Number
(0-127)
Monitor Identification number.
158
M Init
Modem Initialization Sequence
159
Power Mode
[Normal, Ext CO]
Normal: meter uses internal battery
unless greater external voltage is also
applied.
Ext CO: (external cutoff) when exter-
nal power falls below 10.5 V, meter
turns off to prevent battery damage.
161
Timed EXCD Per.
(mm:ss)
Exceedance Time history is recorded
from 10 samples before exceedance
to 10 samples after this period.
162
EXCD Trigger
[Level, Bypass
Bypass: Enables a variable threshold
163
EXCD Time
[Start, Max]
Start: The exceedance will be mea-
sured at the beginning.
Max: The exceedance will be mea-
sured at the maximum.
168
Rt EXCD Report
(No, Yes)
Print exceedance report in real-time.
169
Rt INTV Report
(No, Yes)
Print interval report in real-time.
170
Rt HIST Report
(No, Yes)
Print history report in real-time.