4-2
Setting Parameters Using Function Keys
Model 820 Technical Reference Manual
A complete list of parameters and their assigned
numbers is found in chapter 8. You have already vis-
ited parameter 43, Microphone Polarization, and
found it preset to [200].
Step 2
To access parameters using numeric values, press
m
:
Notice that the flashing (f) has
been replaced by a flashing (n)
indicating the numeric key access.
The flashing parenthesis, indicate
this field is ready to receive
numeric input.
Step 3
To access the Current Time, press
0
and then
6
.
The number 6 is assigned to the Current Time
parameter:
Entering
0
before the new num-
ber will remove any prior parame-
ter settings.
Step 4
Press
m
to prepare the screen for changes in the
Current Time.
Notice the flashing (f) changes to (n)
.
The third way to access Current
Time parameter, press
R
UT
. Remember, press
m
to change or correct the
field and press
e
.