10-2
Continuous and Timer Modes
LxT Manual
TABLE 10-1 shows when measurement histories are created
for each run mode.
TABLE 10-1 Measurement History Creation for Run Modes
Continuous and Timer Modes
For these run modes, when the Measurement History is
enabled, a series of measurements are performed and stored
automatically, each running for a user-defined time interval.
When Measurement History is enabled, the interval time can
be specified, as shown in FIGURE 10-3.
FIGURE 10-2
FIGURE 10-3 Measurement Time Menu
Select the
Time
data field and press
5
. Enter the desired
time and press
5
.
Interval Time Sync
The interval time sync feature ensures that all measurement
records, except the first, begin at a time equal to a multiple
of the measurement time selected. For example, if the
measurement time is five minutes, and the measurement
begins at 08:14:00 (h:m:s format), the first measurement is
cut short such that the subsequent measurements begin at
08:15, 08:20, 08:25, etc.
Run Mode
Measurement History Creation
Manual
when Stop is pressed
Timed Stop
when instrument stops
Stop When Stable
when instrument stops
Continuous
at every pre-specified period (1 minute minimum)
Single Block Timer
at every pre-specified period (1 minute minimum)
Daily Timer
at every pre-specified period (1 minute minimum)
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...