LxT Manual
Technical Specifications
A
-
51
Model PRMLxT2B
The Larson Davis PRMLxT2B is a prepolarized microphone preamplifier for use with Larson Davis
LxT Sound Level Meters. It requires very little supply current and will drive 200 feet of cable. The
preamplifier operates over wide temperature and humidity ranges. It has a built in attenuation of 19 dB
for use with 35.5 mV/Pa sensitivity microphones up to 140 dBSPL.
Specifications
Unless otherwise stated, all electroacoustic values are at 23°C, 50% RH, 7.5 Volt supply, 3 m (10')
cable and equivalent microphone of 18pF.
Frequency response
with respect to the response at 1 kHz with 8.9 Volts rms input and 18 pF equivalent microphone.
2.5 Hz to 10 Hz +0.1, -0.25 dB
10 Hz to 40 kHz +0.1, -0.1 dB
Lower -3 dB limit < 1 Hz
Attenuation
19 dB (typical)
Input Impedance
10 G Ohm // 2.5 pF
Output Impedance
50 Ohm
Maximum Output
2.8 Vpp
143 dB peak for microphones with 35.5 mV/Pa sensitivity
Maximum Output Current
10 mA peak
Distortion
Harmonics <-58 dBc with 0.7 Volt rms output at 1 kHz
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...