C
-
8
Glossary and Formulas
LxT Manual
Dose
See
Noise Dose
.
Dose and Projected Dose
Calculations
Dose is a measure of Sound Exposure and is defined in
ANSI S1.25 Section 4.7 as:
See FIGURE C-1 “DOSE and Pro-
jected DOSE"
where:
D(Q) is the percentage criterion exposure for exchange rate
Q
T
c
is the criterion sound duration
T is the measurement duration in hours
t is the time in hours
L is the SLOW, (or FAST) A-weighted sound level, a
function of time, when the sound level is greater than or
equal to
Lt
, or equals -
∞
when the A-weighted sound level is
less than
Lt
L
t
is the threshold sound level specified by the manufacturer
L
C
is the criterion sound level specified by the manufacturer
Q is the exchange rate in dB, and
q
= the parameter that
determines the exchange rate, where:
q = 10 for a 3dB exchange rate
q = 13.29 = 4/log(2) for a 4dB exchange rate
q = 16.61 = 5/log(2) for a 5dB exchange rate
q = 20 = 6/log(2) for a 6dB exchange rate
D Q
100
T
c
---------
10
L L
c
–
q
--------------
dt
0
T
=
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...