LxT Manual
Glossary and Formulas
C
-
21
T
c
is the criterion sound duration as set by the LxT’s
“Criterion Time Hours” setting which by default is set to 8
hours L
c
is the criterion sound level as set by the LxT’s
“Overall Criterion” or “Current Criterion” settings.
Addition of the term “log(100)” was used to implement the
100 multiplier of the ANSI equation that creates the
percentage. Subtracting the log of the Criterion Time was
used to implement the division of Criterion Time of the
ANSI equation.
Projected Dose in the analyzer is obtained with an equation
similar to that of Dose except that the actual duration (time)
of the measurement is used rather than a Criterion Time, as
thus:
See Standard: ANSI S12.19
where the log(
n
) is the actual time factor,
n
being the total
number of samples taken.
FIGURE C-2 DOSE and Projected DOSE
PROJDOSE=10
log
10
L s
k
----------
s
1
=
n
L
C
k
------
n
log
100
log
+
–
–
%
The currently selected Frequency Weighting (A)
and Detector Response (Slow) for SPL
The currently selected Exchange Rate
Criterion Time and Level
Elapsed measurement time shown as h:mm:ss.s
DOSE
Projected DOSE
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...