LxT Manual
Glossary and Formulas
C
-
25
FIGURE C-3 Sound Exposure Level and Sound Exposure
Sound Exposure Level (SEL,
L
E
)
The total sound energy in a specific time period.
The sound pressure is squared and integrated over a specific
period of time (T
2
-T
1
) this is called the sound exposure and
has the units Pascal squared- seconds or Pascal squared-
hours. P
O
is the reference pressure of 20 µPa and T is the
reference time of 1 second. It is important to note that it is
not an average since the reference time is not the same as the
integration time.
The A-weighted sound exposure level L
AE
is related to a
corresponding measurement of time-average, A-weighted
sound level, L
AT
or L
AEeqT
, by
where
E
A
is the A-weighted sound exposure in pascal-squared
seconds
E
0
is the reference sound exposure of:
Sound Exposure Level
Sound Exposure in hours
Sound Exposure in seconds
dB
T
T
L
dB
E
E
dB
T
p
dt
t
p
L
AT
A
t
t
A
AE
0
0
0
2
0
2
1
2
/
/
/
lg
10
lg
10
lg
10
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...