4-6
Measurement Settings Tabs
LxT Manual
Graph scaling range can be
modified by the user, as described in
“Graph Scale Adjustment” on
page 5-2.
OBA Frequency Weighting
The user can select that the 1/1 and/or 1/3 Octave frequency
analysis modules process data from the A, C or Z weighting
filters.
OBA Max Spectrum Setting
Two methods can be used to define the maximum spectrum:
• At Lmax:
using this method, the maximum values for
each frequency band are those that are being measured at
the instant the overall sound pressure level reached its
maximum value during the measurement period.
• Bin Max:
using this method, the level measured for each
frequency band is the maximum measured during the
measurement period. Since the maximum levels for the
different frequency bands may have occurred at different
times, the ensemble of frequency band maximum levels
may represent a spectrum that never existed at any single
instant during the measurement.
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...