LxT Manual
Measurement Settings Tabs
4-7
Dosimeter 1 and 2 Tabs
Figure 4-6 shows the
Dosimeter 1
tab.
FIGURE 4-6 Dosimeter 1
The
Dosimeter 1
and
Dosimeter 2
tabs are provided to
permit the evaluation of two independent noise dose data
sets. Other than being on separate tabs, they are identical.
Predefined Setups
In most cases, measurements of this type are setup to
conform to specific standards. The LxT permits the user to
create such setups in a single step by simply selecting the
applicable standard. The standards addressed by the LxT and
the corresponding parameters are as shown in Table 4-1
"Predefined Noise Dosimeter Setups".
Standard
Exchange
Rate
Threshold
Criterion
Level
Hours
OSHA-1
5
90
90
8
OSHA-2
5
80
90
8
ACGIH
3
80
85
8
NIOSH
3
80
85
8
IEC
3
Not Enabled 85
8
Table 4-1 Predefined Noise Dosimeter Setups
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...