LxT Manual
Calibration
7-1
C H A P T E R
7
Calibration
This chapter describes both the purposes and steps for
calibrating the SoundTrack LxT
and SoundExpert
LxT.
Calibration Overview
Sensitivity Determination
The primary role of sound level meter calibration is to
establish a numerical relationship between the sound level at
the diaphragm of the microphone and the voltage measured
by the meter so that the sound pressure level can be read
directly from the display of the meter in units of dB. The
result of a calibration is the determination of the sensitivity
of the meter, including microphone and preamplifier,
typically in units of dB re 1V/Pa or mV/Pa.
Overload/Under Range Conditions
A secondary role of calibration is to determine the sound
level that would overload the instrument and the minimum
sound level that can be accurately measured, referred to as
the under range level. This requires a knowledge of the
electrical noise levels of the microphone, preamplifier, and
the instrument circuitry.
Calibration Stability
The LxT should maintain a stable value of sensitivity over
long periods of time. Significant changes in sensitivity, or a
pattern of small but regular sensitivity changes, are
indicative of problems with the measurement system, calling
for laboratory calibration and possibly service. To assist the
user in identifying these situations, the LxT provides two
notifications:
Calibration History
Data and date/time of the most recent ten calibrations.
Summary of Contents for SoundExpert LxT
Page 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Page 55: ...4 18 Measurement Settings Tabs LxT Manual...
Page 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Page 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Page 110: ...LxT Manual Certification 7 17...
Page 143: ...11 6 Control Panel Data Explorer LxT Manual...
Page 167: ...12 24 Options LxT Manual...
Page 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Page 185: ...15 6 File System LxT Manual...
Page 191: ...16 6 Miscellaneous Parameters LxT Manual...
Page 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...