TROUBLESHOOTING AND CUSTOMER SUPPORT 

Verify Functionality 

Pressing the power button should activate the LCD and provide a real-time irradiance reading. Direct the sensor 
head toward a light source and verify the irradiance reading responds. Increase and decrease the distance from 
the sensor to the light source to verify that the reading changes proportionally (decreasing irradiance with 
increasing distance and increasing irradiance with decreasing distance). Blocking all radiation from the sensor 
should force the reading to zero. 

Battery Life 

When the meter is maintained properly the coin cell battery (CR2320) should last for many months, even after 
continuous use. The low battery indicator will appear in the upper left hand corner of the LCD display when the 
battery voltage drops below 2.8 V DC. The meter will still function correctly for some time, but once the battery is 
drained the pushbuttons will no longer respond and any logged measurements will be lost. 

Pressing the power button to turn off the meter will actually put it in sleep mode, where there is still a slight 
amount of current draw. This is necessary to maintain the logged measurements in memory. Therefore, it is 
recommended to remove the battery when storing the meter for many months at a time, in order to preserve 
battery life.  

Low-Battery Error after Battery Replacement 

A master reset will usually correct this error, please see the master reset section for details and cautions. If a 
master reset does not remove the low battery indicator, please double check that the voltage of your new batter is 
above 2.8 V, this is the threshold for the indicator to turn on. 

Master Reset 

If a meter ever becomes non-responsive or experiences anomalies, such as a low battery indicator even after 
replacing the old battery, a master reset can be performed that may correct the problem. Note that a master reset 
will erase all logged measurements from memory.  

Step 1:  press the power button so that the LCD display is activated. 

 

Step 2: Slide the battery out of the holder, which will cause the LCD display to fade out. 

Step 3: After a few seconds, slide the battery back into the holder. 

The LCD display will flash all segments and then show a revision number (e.g. “R1.0”). This indicates the master 
reset was performed and the display should return to normal. 

Larson Electronics, LLC

 

  Phone: (800) 369-6671 

 Fax: (903) 498-3364 

  www.larsonelectronics.com 

Summary of Contents for XLE-HHLM-VISNIR-360NM.1200NM-V1

Page 1: ...OWNER S MANUAL SOLAR RADIATION METER Part XLE HHLM VISNIR 360NM 1200NM V1 Larson Electronics LLC Phone 800 369 6671 Fax 903 498 3364 www larsonelectronics com...

Page 2: ...rnational Organization of Standardization ISO second class and first class classifications but because of limited spectral sensitivity they do not meet the spectral specification necessary for WMO or...

Page 3: ...rtainty 5 see Calibration Traceability below Measurement Repeatability Less than 1 Long term Drift Non stability Less than 2 per year Non linearity Less than 1 up to 1750 W m 2 Response Time Less than...

Page 4: ...a for the photodiode were obtained from the manufacturer Mean temperature response of four Larson silicon cell pyranometers Temperature response measurements were made at approximately 10 C intervals...

Page 5: ...rado Cosine response was calculated as the relative difference of pyranometer sensitivity at each solar zenith angle to sensitivity at 45 solar zenith angle The blue symbols are AM measurements the re...

Page 6: ...ogged measurements on the LCD display Logging To choose between manual or automatic logging push the mode button once and use the up down buttons to make the appropriate selection SMPL or LOG Once the...

Page 7: ...fers in clear sky and cloudy conditions thus measurements in cloudy conditions result in spectral error because sensors are calibrated in clear sky conditions Spectral response of Larson series pyrano...

Page 8: ...m Agricultural Soil 12 1 Reflected from Forest Soil 4 1 Reflected from Desert Soil 3 0 Reflected from Water 6 6 Reflected from Ice 0 3 Reflected from Snow 13 7 Spectral error for Larson series pyranom...

Page 9: ...acy the sky must be completely clear as reflected radiation from clouds causes incoming radiation to increase above the value predicted by the clear sky calculator Measured values of total shortwave r...

Page 10: ...nt draw This is necessary to maintain the logged measurements in memory Therefore it is recommended to remove the battery when storing the meter for many months at a time in order to preserve battery...

Page 11: ...PU voltage below minimum Fix replace CR2320 battery and perform master reset Larson Electronics LLC Phone 800 369 6671 Fax 903 498 3364 www larsonelectronics com THESE INSTRUCTIONS MAY NOT COVER ALL D...