LED FLASHING MODES

EL-USB-1 features 2 LEDs.

• 

The first LED flashes green or red to indicate alarm level status.

• 

The second LED flashes orange to indicate a problem condition on the EL-USB-1.

It will flash red when the logged temperature has exceeded a Low or High alarm level. By default hold is disabled. In this mode the 
red LED will no longer continue to flash after the logged temperature has returned to normal from an alarm condition.

Hold can be turned on via the control software. In this mode the red LED that indicates an alarm condition will continue to flash, even 

after the logged temperature has returned to normal. The red LED will effectively have latched into its alarm condition. This feature 

ensures that the user is notified that an alarm level has been exceeded, without the need to download the data from the logger. 

Green double flash              

The data logger is not currently logging, but is primed to start at a later date and time (delayed start).

Green single flash

The data logger is currently logging. No alarm.

Green and Orange single flash

The data logger is full and has stopped logging. No alarm.

Red single flash

The data logger is currently logging. Low alarm.

Red and Orange single flash

The data logger is full and has stopped logging. Low alarm.

Red double flash

The data logger is currently logging. High alarm.

Red and Orange double flash

The data logger is full and has stopped logging. High alarm.

Orange single flash

The battery is running low as its voltage has dropped below 3.3V. Data logging will continue until the battery 
voltage drops below 2.8V.

No LEDs flash

The data logger is stopped, the battery is empty or there is no battery

Red/Green 

LED

Orange 
LED

EL-USB-1          

Temperature Data Logger with USB Interface

www.lascarelectronics.com 

    

         Specifications liable to change without prior warning   

      Issue 12         07/2012                     S.L                  Applies to EL-USB-1              Page 2 of 4

Summary of Contents for EL-USB-1

Page 1: ...indows control software Environmental protection to IP67 ORDERING INFORMATION Standard Data Logger Data Logger Software on CD Battery Replacement Battery EL USB 1 BAT 3V6 1 2AA Specifications liable to change without prior warning Issue 12 07 2012 S L Applies to EL USB 1 Page 1 of 4 www lascarelectronics com At 25 C and 1 m logging rate Specifications Minimum Typical Maximum Unit Measurement range...

Page 2: ...r Green double flash The data logger is not currently logging but is primed to start at a later date and time delayed start Green single flash The data logger is currently logging No alarm Green and Orange single flash The data logger is full and has stopped logging No alarm Red single flash The data logger is currently logging Low alarm Red and Orange single flash The data logger is full and has ...

Page 3: ...th your supplier that the battery you are ordering is press fit and is not fitted with solder tags Before replacing the battery remove the EL USB 1 from the PC Note Leaving the EL USB 1 plugged into the USB port for longer than necessary will cause some of the battery capacity to be lost WARNING Handle lithium batteries carefully observe warnings on battery casing Dispose of in accordance with loc...

Page 4: ...cy temperature humidity dew point 35 to 80 C 31 to 176 F 0 to 100 RH 0 3ºC 0 6ºF 2 0 RH 1 5ºC 3ºF 4 0 RH 16 382 3 6V AA 1 year EL USB 2 LCD Temperature humidity dew point with LCD 35 to 80 C 31 to 176 F 0 to 100 RH 0 5ºC 1ºF 3 0 RH 2ºC 4ºF 6 0 RH 16 379 3 6V AA 1 year EL USB 2 LCD Increased accuracy tem perature humidity dew point with LCD 35 to 80 C 31 to 176 F 0 to 100 RH 0 3ºC 0 6ºF 2 0 RH 1 5º...

Reviews: