26
a. Ensure both pivot bearings are in instrument plate, under pivot end of
sensor.
b. Position the tension spring under the calibration end of the sensor.
3. Insert the 8-32 Phillips head screw with washers into pivot end (refer to
drawing).
a. Tighten the screw until bottomed out and back off 1/8 turn.
4. Insert the x-axis calibration screw through the instrument plate and
tension spring.
a. Thread the cal. screw into the brass calibration bar until approximately
1/4 inch of it is through.
5. Push the sensor standoffs against the sensor assembly and tighten the
guide screws.
6. Replace base plate assembly (see procedure #104)
7. Adjust the cross axis grade alarm dead band (see procedure #108)
#107
Circuit Board Removal and Replacement
Tools required:
Grounded soldering iron
Time:
.5
FIG. 9 Circuit Board Connections and Adjustments
Connections for
Slope Sensor
Connections for
Cross Sensor
Laser Power Pot.
Grade Alarm Pots.
Microprocessor
Grade Alarm Potentiometers
S-Axis
X-Axis
Clockwise adjustment
makes deadband
smaller
Summary of Contents for pro shot L1-M
Page 1: ......
Page 3: ......
Page 6: ...3 External Parts Detail Model L1 M Fig 1 ...
Page 7: ...4 External Parts Detail Model L1 S Fig 1A ...
Page 8: ...5 Internal Parts Detail Model L1 M Fig 2 10 0495 LCD assembly ...
Page 9: ...6 Internal Parts Detail Model L1 S Fig 2a ...
Page 10: ...7 Housing Lighthouse Model L1 S ...
Page 11: ...8 Motor Problems Model L1 M Only ...
Page 12: ...9 Motor Problems Model L1 S Only ...
Page 13: ...10 Power Problems ...
Page 15: ...12 Calibration Problems ...
Page 16: ...13 Calibration Problems 2 ...
Page 17: ...14 Calibration Problems 3 ...
Page 18: ...15 Low Battery Indication Not cured with new Batteries ...
Page 38: ...35 PCB Layout ...
Page 39: ...36 Schematic for L1 M S ...
Page 41: ...38 ...