Duraplane 360
17
EN
Due to the special optics required to generate a continuous 360° laser
line, the underlying technology may cause differences in brightness in
different areas of the line. This may lead to different ranges in hand
receiver mode.
!
A1
A2
2.
1.
A3
A2
A2
A1
4.
3.
A3
A2
<
0,2 mm / m = OK
Preparing the calibration check
It is possible for you to check the calibration of the laser. To do this, position
the device
midway
between 2 walls, which must be at least 5 metres apart.
Switch the device on
(Laser line on)
. The best calibration results are
achieved if the device is mounted on a tripod.
1.
Mark point A1 on the wall.
2.
Turn the device through 180° and mark point A2. You now
have a horizontal reference between points A1 and A2.
Performing the calibration check
3.
Position the device as near as possible to the wall
at the height of point A1.
4.
Turn the device through 180° and mark point A3. The
difference between points A2 and A3 is the tolerance.
When A2 and A3 are more than 0.2 mm / m apart, an adjustment
is necessary. Contact your authorised dealer or else the UMAREX-
LASERLINER Service Department.
!