Preparing the calibration check:

It is possible for you to check the calibration of the laser. To do this, position 
the device 

midway

 between 2 walls, which must be at least 5 metres apart. 

Switch the device on (Laser cross ON). The best calibration results are achie-
ved if the device is mounted on a tripod.  

1.

 Mark point A1 on the wall.

2. 

Turn the device 

through 180° and 
mark point A2.
You now have a 
horizontal reference 
between points A1 
and A2.

Checking the vertical line: 

Position the device about 5 m from a wall. Fix a 

plumb bob with a line of 2.5 m length on the wall, making sure that the bob 
can swing freely. Switch on the device and align the vertical laser to the plumb 
line. The precision is within the specified tolerance if the deviation between the 
laser line and the plumb line is not greater than ± 1.5 mm.

Checking the horizontal line:

Position the device about 5 m from a wall and 
switch on the cross laser. Mark point B on the wall. 
Turn the laser cross approx. 2.5 m to the right and 
mark point C. Check whether the horizontal line 
from point C is level with point B to within ± 2 mm. Repeat the process by 
turning the laser to the left.

Performing the calibration check:

3.

 Position the device as near as possible to the wall at the height of

point  A1.

4.

 Turn the device 

through 180° and 
mark point A3. The 
difference between 
points A2 and A3 is 
the tolerance.

A1

A2

2.

1.

i

Note:

 If points A2 and A3 are more than 2 mm / 10 m, the device is in 

need of calibration. Contact your authorised dealer or else the UMAREX-
LASERLINER Service Department.

B

C

2,5 m

< 2 mm = OK

A3

A2

A2

A1

4.

3.

A3

A2

2 mm / 10 m = OK

SuperCross-Green 3

11