ThermoSpot Pro
7
1x
8
2x
1.
2.
E 0,01...0,99
...
3.
07
GB
A built-in sensor head detects the material/surface-speci
Ö
c infrared rays emitted by every object. The level of
these emissions is determined by the material‘s emission coef
Ö
cient (0.01 to 0.99). This instrument is preset
to an emission coef
Ö
cient of 0.95, which is applicable to most organic materials as well as plastics, ceramics,
wood, rubber and stone. For details of those materials with a different emission coef
Ö
cient, please refer to
the table under point 9 on the next page.
Unknown emission coef
Ƃ
cient:
Apply masking tape or black matt paint to the surface of the area whose temperature you wish to measu-
re. Wait until the tape/paint has heated up. The temperature of the surface can then be measured with an
emission coef
Ö
cient of 0.95.
Setting the emission coef
Ƃ
cient
The MAX display determines the maximum measured value
within the context of a continuous measurement (keep
trigger button pressed). As soon as you release the trigger
button and press it again to start a new measurement, the
MAX value will be reset and the instrument will start the
process of recording the maximum measured value all over
again.
MAX display
Laser circle:
The laser circle is a targeting aid to sight the location for the infrared measurement. Temperature measure-
ment is only performed on the surface within the laser circle. Please ensure that the space between the
instrument and surface to be measured is free of disturbances (steam, gas, contamination, glass).