22 of 49
June 2012
Diagram 4: Trigger Operation
Set ‘TRG’ mode:
Pr
ess 5 so that ‘Trg’ is displayed in the bottom corner of the display. If you
have just entered waveform mode or have just changed range, press
to
reset the trigger level to the middle of the screen. The trigger level is shown
as a horizontal line across the screen.
Use the up and down keys to adjust the vertical trigger point until a stable
waveform is achieved
and ‘Trig’ is shown at the top of the display.
If the y scale is set from a decade setting to a more sensitive ‘5’ or ‘2’ setting,
the trigger level will hold the same absolute value so that the same actual
trigger point is maintained. This may mean that it will disappear off the top of
the screen. If you want to relocate the trigger level into the visible area, press
and the trigger level will be reset to the middle of the screen.
Use the left and right buttons to move the trigger point horizontally. The
horizontal trigger point is shown by a small ‘T’ at the top of the graph.
Press 5 to turn off the trigger mode and the horizontal trigger level line is no
longer shown. However the vertical trigger point is indicated by a small index
mark to the left of the grid.