BL57-NE
(E)
3-7
Mounting surface
Datum surface
3
Check the mounting surface and the datum surface of the scale for dust and dirt. If the surfaces are dirty,
wipe them clean with gauze moistened with alcohol.
Fig. 3-7
4
Insert the scale between the scale clamp and the mounting surface while pressing the datum surface of
the scale (see Fig. 3-5) lightly against the reference abutment surface (Note). Gradually and alternately
tighten screws A and B of the scale clamp indicated in the figure. (Use only the left-side scale clamp for
clamps with two screws, and use the center scale clamp for clamps with three screws.) (Final fastening
torque: 1.4 N
·
m)
Fig. 3-8
B
A
B
A
B
A
BL57-003NE
BL57-006NE, 007NE, 026NE, 027NE, 032NE, 046NE
BL57-012NE, 016NE, 017NE, 022NE, 036NE, 037NE, 042NE
B
A
B
A
BL57-056NE, 066NE, 076NE, 086NE, 096NE, 106NE