Aqualine
15.03.18/ YACE0086
Brief operating instructions
9
2 Brief operating instructions
This brief instruction shall give you the possibility to operate the unit quickly. For safe oper-
ation of the unit it is absolutely necessary to read carefully all the instructions and safety
notes!
1. Assemble unit and add items as appropriate (
6).
2. Fill the unit with decalcified water (
6.3). Take care of the level of the bath liquid! (
6.2).
The units are designed for operation with non-flammable liquids to DIN EN 61010-2-010.
3. Connect the unit only to a socket with a protective earth (PE) connection.
Compare the information on the rating label with the supply details.
4. Switch on at the mains switch
.
5. The controller runs a self-test after switching on. For a few seconds all segments are flashing.
6. Setting the set point temperature
First the current bath temperature is displayed.
Press the key
shortly. The display shows
SEt
. Press
shortly again
the set point
will be displayed.
Adjust the set-point using the keys
or
.
After 30 seconds without entry the current bath
temperature will be displayed again.
Save the set value with twice
or wait for
30 s till the set point value will be saved
automatically.
2§1
°
C
!
fnc
set