7.3 Temperature set-point setting
Display and modification of the operating temperature set-point:
Press the key
shortly. The display shows
SEt
. Press
shortly again
the set point will be displayed.
Adjust the set-point using the keys
or
.
After 30 seconds without entry the bath tem-
perature will be displayed again.
Save the set value twice with
or wait for
30 s till the set point value will be saved
automatically.
The LED 2 is illuminated, if the heater feeds
heat to the bath.
7.4 Controller error signals
LED
display
Visual alarm switch-
off
Device status
Cause
Remedy
E1
Automatic reset when
the signal returns to the
acquisition range.
Heater is OFF.
No or to less water in the
tank.
Topping up with de-
calcified water.
Temperature above
110 °C. Temperature sen-
sor, connecting wires or
controller is defective.
Contact LAUDA Ser-
vice Constant Tem-
perature Equipment.
3/3
°