Technical data

 

23

 

 

10 Technical Data 

The figures have been determined according to DIN 12876. 

Aqualine water baths 

 

AL 2 

AL 5 

AL 12 

AL 18 

AL 25 

Operating temperature range 

°C 

25 

– 95 

Temperature stability at 37 °C 

±0.2 

Resolution setting 

K

 

0.1 

Absolute display accuracy 

±2 

Heater power @ 

 

 

 

 

 

 

230 V; 50/60 Hz  

kW 

0.45 

0.45  

1.0 

1.2 

1.2  

115 V; 60 Hz  

kW 

0.45 

0.45 

1.0 

1.2 

1.2 

100 V; 50/60 Hz 

kW 

0.34 

0.34 

0.76 

0.9 

0.9 

Heater surface loading @ 
230 V 

W/cm²

 

2.4 

2.4 

2.24 

1.38 

1.38 

Power consumption @ 

 

 

 

 

 

 

230 V; 50/60 Hz  

kW 

0.5 

0.5 

1.0 

1.2 

1.2 

115 V; 60 Hz 

kW 

0.5 

0.5 

1.0 

1.2 

1.2 

100 V; 50/60 Hz 

kW 

0.4 

0.4 

0.8 

0.91 

0.91 

least mains fuse 
protection @ 

230 V  
115 V  
100 V 

A  
A  

3  
5  

3  
5  

6  

10  

7  

12  
11 

7  

12  
11 

Max. bath volume 

0.9 

– 1.7 

– 5 

– 12 

– 18 

– 25 

Bath vessel 

--- 

Deep-drawn inner tank in stainless steel 1.4301  

conforming to SAE 30304 AISI 304. 

Bath opening (w x d) 

mm 

300 x 151 

329 x 300 

505 x 300 

Bath depth 

mm 

65 

 

150 

 

200 

Footprint (w x d) 

mm 

347 x 183 

375 x 335 

550 x 335 

Height of top edge of bath with-
out cover 

mm 

 

234 

 

284 

Height incl. gable cover 

mm 

311 

333 

383 

Weight 

kg 

4.5 

6.8 

9.8 

Housing 

--- 

Powder-coated steel sheet. 

Ambient temperature range 

°C 

– 40 

Humidity 

--- 

Maximum relative humidity 80 % for temperatures up to 31 °C, de-

creasing linearly to 50 % relative humidity at 40 °C. 

Storage temperature range 

°C 

-20 

– 60 

Summary of Contents for AL 2

Page 1: ...2SHUDWLQJ LQVWUXFWLRQV 8 TXDOLQH...

Page 2: ......

Page 3: ...arrstra e 41 43 97922 Lauda K nigshofen Germany replaces release 04 2016 b 06 2012 05 09 03 09 03 07 YACE0086 Telephone 49 0 9343 503 0 Fax 49 0 9343 503 222 E mail info lauda de Internet http www lau...

Page 4: ......

Page 5: ...ter Never operate the equipment without sufficient water level Do not start up the equipment if it is damaged or leaking the supply cable is damaged Switch off the equipment and pull out the mains plu...

Page 6: ...7 1 MAINS CONNECTION 17 7 2 SWITCHING ON 17 7 3 TEMPERATURE SET POINT SETTING 18 7 4 CONTROLLER ERROR SIGNALS 18 8 MAINTENANCE 19 8 1 LOW LEVEL PROTECTION 19 8 2 UNLOCK THE SAFETY TEMPERATURE LIMITER...

Page 7: ...ns Class B in accordance with CISPR 11 Worldwide No limitation Instructions for Class A digital devices USA This equipment has been tested and found to comply with the limits for Class A digital devic...

Page 8: ...cket Parts of the bath cover may reach surface temperatures above 70 C when operating at higher tem peratures Take care when touching it Always pull out the mains plug before cleaning maintenance or m...

Page 9: ...010 2 010 3 Connect the unit only to a socket with a protective earth PE connection Compare the information on the rating label with the supply details 4 Switch on at the mains switch 5 The controller...

Page 10: ...and functional elements 1 Gable cover made of plastic material swiveling and demountable 2 Temperature controller with LED display 3 Mounting feet 4 Control keys 5 Status LED K1 and K2 6 LED display 7...

Page 11: ...Aqualine 15 03 18 YACE0086 Control and functional elements 11 10 Name plate and resetting button for the safety temperature limiter Aqualine AL 5 Aqualine AL 12 Aqualine AL 18 10...

Page 12: ...terial Polycarbonate The bath cover made of steel is not interchangeable against the bath cover made of plastic material 4 4 Temperature display controller and safety circuit The bath tank is heated f...

Page 13: ...number Quantity Designation HGH 194 1 Bath gable cover made of transparent plastic material Polycarbonate fitted for every Aqualine water baths AL 2 and AL 5 HGH 195 1 Bath gable cover made of transp...

Page 14: ...ation opening on the bottom of the unit and the heatsink at the back of the unit Keep a clear distance of at least 20 cm The gable cover has two positions Closed or opened In the position shown see fi...

Page 15: ...mm minimum level is possible Emptying Switch off the water bath and withdraw the mains plug Empty the bath with the aid of a suction pump or other extracting device Siphon Hand Pump LAUDA catalogue nu...

Page 16: ...um car bonate liter of water TIP Water baths can be operated ideally with LAUDA Aqua 90 available in container sizes of 5 10 or 20 liters order number LZB 120 LZB 220 or LZB 320 There is a risk of ele...

Page 17: ...6 2 Note for electric installation on site The devices must be protected with a 16 ampere circuit breaker fitted during installation Exception Devices with 13 ampere UK plugs 7 2 Switching on Switch...

Page 18: ...e set value twice with or wait for 30 s till the set point value will be saved automatically The LED 2 is illuminated if the heater feeds heat to the bath 7 4 Controller error signals LED display Visu...

Page 19: ...t is a low water level in the tank 7 4 Above approx 150 C bottom temperature a mechanical safety temperature limiter cuts off the complete voltage supply Thereupon the LED display goes out This protec...

Page 20: ...uses as specified 2 off T slow blow 12 5 A EES 014 1 off T slow blow 0 2 A EES 069 Size 6 3 x 32 fuses located in the unit Changing the fuses Before opening withdraw the mains plug pick up the gable c...

Page 21: ...om the name plate This avoids queries and supply of incorrect items Name plate Your contact for service and support LAUDA Service Constant Temperature Equipment Telephone 49 0 9343 503 350 English and...

Page 22: ...AL 25 UE 029 Rack RN 18 1 AL 12 2 x AL 18 AL 25 UE 030 Rack RN 18 2 AL 12 2 x AL 18 AL 25 UE 031 Rack RN 30 AL 12 2 x AL 18 AL 25 UE 032 Rack RQ 13 AL 5 3 x AL 12 5 x AL 18 AL 25 UG 105 Rack RQ 18 1 A...

Page 23: ...kW 0 5 0 5 1 0 1 2 1 2 100 V 50 60 Hz kW 0 4 0 4 0 8 0 91 0 91 least mains fuse protection 230 V 115 V 100 V A A A 3 5 4 3 5 4 6 10 9 7 12 11 7 12 11 Max bath volume L 0 9 1 7 1 5 2 12 3 18 3 25 Bath...

Page 24: ...tion class according to DIN EN 61140 VDE 0140 1 Protection class I EU conformity The device complies with the basic health and safety requirements outline in the Directives listed below Low voltage Di...

Page 25: ...il LAUDA voir plaque signal tique Typ Type Type Serien Nr Serial no No de s rie mit folgendem Medium betrieben wurde was used with the below mentioned media a t utilis avec le liquide suivant Dar ber...

Page 26: ......

Page 27: ......

Page 28: ...LAUDA DR R WOBSER GMBH CO KG P O Box 12 51 97912 Lauda Koenigshofen Germany Phone 49 9343 503 0 Fax 49 9343 503 222 E mail info lauda de Internet www lauda de...

Reviews: