Generally, this terminating network is integrated on the PC plug-in card (RS 485).
Note the following aspects:
–
The interface operates with one stop bit, no parity bit and with eight data bits.
–
Transfer speed alternatively: 2400, 4800, 9600 (factor setting) or 19200 baud.
–
The device address always precedes the RS 485 commands. Up to 127 addresses are possible. The ad-
dress must always consist of three figures
(A000_... to A127_...).
–
The command from the computer must be terminated with a CR.
–
The response from the thermostat is always terminated with a CR.
CR = Carriage Return (Hex: 0D)
Example:
Set-value transfer of 30.5 °C to the thermostat with address 15.
Computer
Thermostat
„A015_OUT_SP_00_30.5“CR
„A015_OK“CR
LiBus module (catalogue number LRZ 920) has a socket (70S) for con-
nection of components via the LAUDA device bus LiBus (Command re-
mote control, shut down/reverse flow protection, cooling water valve).
LiBus = LAUDA internal device bus (CAN based)
For extension cord for LiBus, see accessories (
F.
3
.4
RS 485 protocol
F
.
4
LiBus module
Summary of Contents for ECO Silver E 10 S
Page 1: ...ECO Silver Operation manual ...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i ...
Page 79: ...10 2019 i ECO Silver 79 124 ...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings ...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish ...
Page 113: ...10 2019 i ECO Silver 113 124 ...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page ...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page ...
Page 122: ...122 124 ECO Silver 10 2019 i ...
Page 125: ......
Page 126: ......
Page 127: ......