116 / 124
ECO Silver
Command
Meaning
IN_DO_01
Status of Contact Output 1:
0 = NO contact open/ 1 = NO contact closed.
IN_DO_02
Status of Contact Output 2:
0 = NO contact open/ 1 = NO contact closed.
IN_DO_03
Status of Contact Output 3:
0 = NO contact open/ 1 = NO contact closed.
IN_MODE_00
Keypad: 0 = released / 1 = locked.
IN_MODE_01
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analog / 3 = ext. Serial.
IN_MODE_02
Standby operation: 0 = Device ON / 1 = Device OFF.
IN_MODE_03
Keypad remote control unit Command: 0 = released / 1 = locked.
IN_MODE_04
Setpoint offset source: 0 = normal / 1 = ext. Pt / 2 = ext. Analog / 3 = ext. Serial.
TYPE
Query of the device type (response = "ECO")
VERSION_R
Query of the software version number of the control system.
VERSION_S
Query of the software version number of the protection system.
VERSION_B
Query of the software version number of the Command remote control.
VERSION_T
Query of the software version number of the cooling system.
VERSION_A
Query of the software version number of the analog module.
VERSION_V
Query of the software version number of the RS 232/485 module.
VERSION_D
Query of the software version number of the digital module.
VERSION_M_0
Query of the software version number of the solenoid valve (cooling water).
VERSION_M_3
Query of the software version number of the solenoid valve (shut-off valve 1).
VERSION_M_4
Query of the software version number of the solenoid valve (shut-off valve 2).
VERSION_M_5
Query of the software version number of the high temperature cooler.
VERSION_E
Query of the software version number of the external Pt100 module.
STATUS
Query of the device status 0 = OK, -1 = Error.
STAT
Query of the error diagnosis response:
XXXXXXX
X = 0 no error, X = 1 error
1st character = Error
2nd character = Alarm
3rd character = Warning
4th character = Overtemperature
5th character = Low Level
6th character = 0
7th character = External control value missing
RMP_IN_00_XXX
Query of a program segment XXX
(Response: e.g. 030.00_00010.00_005.00_001.00 => Set-point temperature =
30.00 °C, Time = 10 min, Tolerance = 5.00 K, Pump stage = 1).
RMP_IN_01
Query of the current segment number.
RMP_IN_02
Query of the set program loops.
RMP_IN_03
Query of the current program loops.
RMP_IN_04
Query of to which program further commands refer.
RMP_IN_05
Query of which program is currently running (0 = none).
Summary of Contents for ECO Silver E 10 S
Page 1: ...ECO Silver Operation manual ...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i ...
Page 79: ...10 2019 i ECO Silver 79 124 ...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings ...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish ...
Page 113: ...10 2019 i ECO Silver 113 124 ...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page ...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page ...
Page 122: ...122 124 ECO Silver 10 2019 i ...
Page 125: ......
Page 126: ......
Page 127: ......