List of "Alarm and warning codes"
Alarms
Alarm code
Meaning
1
Low Level Pump
Pump runs too fast (low level)
2
Low Level Pump
Low level in the float
3
Overtemperature
Overtemperature (T > Tmax)
4
Pump blocked
Pump blocked (standstill)
5
Connection Command
Remote control unit command triggered in running operation
9
T ext Pt100
External Pt100 actual value is not present.
10
T ext analog
External analog actual value is not present.
11
T ext serial
External serial actual value is not present.
12
Input Analog 1
Analog module: Current interface 1, interruption.
13
Input Analog 2
Analog module: Current interface 2, interruption.
15
Digital Input
Error on digital input
Warnings
Code 0XX Control system
Meaning
Code 3XX SmartCool
Meaning
1
CAN receive overf
Overflow during CAN reception
1
CAN receive overf
Overflow during CAN reception
2
Watchdog Reset
Watchdog reset
2
Watchdog Reset
Watchdog reset
3
T_il limit active
til llmit active
3
adaption missing
No adaption run
4
T_ih limit active
tih llmit active
4
Pressure switch
activated
Pressure Switch in cooling circuit trig-
gered
5
corrupt parameter
Inadmissible internal parameter
5
Clean condenser
Clean condenser
6
corrupt progr
Inadmissible programmer data
6
TO1 range Klixon
Injection temperature outside value
range
7
Invalid Parameter
Inadmissible parameter in memory
7
Invalid Parameter
Inadmissible parameter in memory
8
CAN system
Problem during internal data interchange
8
CAN system
Problem during internal data interchange
9
Unknown Modul
Unknown module connected
9
Unknown Modul
Unknown module connected
10
SW control too old
Software version of control panel too old
10
SW control too old
Software version of control panel too old
11
SW safety too old
Software version of protection too old
11
SW safety too old
Software version of protection too old
12
SW command too old Software version of command remote
control unit too old
12
SW command too old Software version of command remote
control unit too old
13
SW cool too old
Software version of cooling module too
old
13
SW cool too old
Software version of cooling module too
old
14
SW analog too old
Software version of analog too old
14
SW analog too old
Software version of analog too old
15
SW serial too old
Software version of RS 232 too old
15
SW serial too old
Software version of RS 232 too old
16
SW contact old
Software version of contact module too
old
16
SW contact old
Software version of contact module too
old
17
SW Valve 0 old
Software version of solenoid valve 0 too
old
17
SW Valve 0 old
Software version of solenoid valve 0 too
old
18
SW Valve 1 old
Software version of solenoid valve 1 too
old
18
SW Valve 1 old
Software version of solenoid valve 1 too
old
19
SW Valve 2 old
Software version of solenoid valve 2 too
old
19
SW Valve 2 old
Software version of solenoid valve 2 too
old
20
SW Valve 3 old
Software version of solenoid valve 3 too
old
20
SW Valve 3 old
Software version of solenoid valve 3 too
old
21
SW Valve 4 old
Software version of solenoid valve 4 too
old
21
SW Valve 4 old
Software version of solenoid valve 4 too
old
26
SW HTC old
Software version of high temperature
cooler too old
26
SW HTC old
Software version of high temperature
cooler too old
27
SW Ext Pt100 old
Software version of external Pt100 too
old
27
SW Ext Pt100 old
Software version of external Pt100 too
old
Summary of Contents for ECO Silver E 10 S
Page 1: ...ECO Silver Operation manual ...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i ...
Page 79: ...10 2019 i ECO Silver 79 124 ...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings ...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish ...
Page 113: ...10 2019 i ECO Silver 113 124 ...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page ...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page ...
Page 122: ...122 124 ECO Silver 10 2019 i ...
Page 125: ......
Page 126: ......
Page 127: ......