54 / 124
ECO Silver
When the bath temperature is located above the overtemperature switch-
off point, a two-tone alarm sounds. "Overtemperature" appears in the dis-
play, the heater switches off on all poles and the pump and chiller are
switched off via the electronics.
–
Rectify the cause of the malfunction.
–
Wait until the bath temperature has cooled below the over-
temperature switch-off point or set the overtemperature switch-
off point higher than the bath temperature.
If "Overtemperature" appears in the display:
–
Unlock the "Overtemperature" display with .
Before a longer unsupervised operation check the overtemperature protection:
–
Slowly reduce T
max
as described in (
7.4.5). The thermostat should switch off when the actual temperature is
greater than T
max
.
An alarm message (step 1-2, see above) should follow.
–
Reset the switch-off point to be higher than the bath temperature.
–
Unlock the "Overtemperature" display with .
8.1.2
Low level: Alarm and checking
When the liquid level falls so far that the heaters are no longer completely
covered with liquid, a two-tone alarm sounds. "Low Level Pump" appears in
the display, the heater switches off on all poles and the pump and chiller are
switched off via the electronics.
–
Rectify the cause of the malfunction.
–
Top up the missing heat transfer liquid (
–
Unlock the "Low Level Pump" display with .
Check the safety system at regular intervals (
8.3.2) by lowering the bath level. Do not carry out this test at a bath
temperature below 0 °C or above 50 °C in order to avoid dangers due to temperatures that are too hot or too cold.
An alarm message (step 1-2, see above) should follow.
–
Top up with heat transfer liquid.
–
Unlock the "Low Level Pump" display with .
Switch the device off immediately and withdraw the mains plug if irregularities occur when checking the safety devices.
Contact LAUDA Service Constant Temperature Equipment (
8.7) or your local service.
No. Alarm
Overtemperature
3
Overtemperature
DISPLAY
OK
STOP
No. Alarm
Low Level Pump
1
Low Level Pump
DISPLAY
OK
STOP
Summary of Contents for ECO Silver E 10 S
Page 1: ...ECO Silver Operation manual ...
Page 2: ......
Page 4: ......
Page 78: ...78 124 ECO Silver 10 2019 i ...
Page 79: ...10 2019 i ECO Silver 79 124 ...
Page 80: ...80 124 ECO Silver 10 2019 i Appendix with settings ...
Page 111: ...10 2019 i ECO Silver 111 124 5 Click on Continue installation 6 Click on the key Finish ...
Page 113: ...10 2019 i ECO Silver 113 124 ...
Page 120: ...120 124 ECO Silver 10 2019 i Blank page ...
Page 121: ...10 2019 i ECO Silver 121 124 Blank page ...
Page 122: ...122 124 ECO Silver 10 2019 i ...
Page 125: ......
Page 126: ......
Page 127: ......