Starting up
40
☞
−
When the switch-off point is being adjusted by more than 2°C
!
display shows MAX
and actual overtemperature switch-off point with 1°C resolution for approx. 4 s.
−
The position of the potentiometer is decisive for the setting. The display is just a help
for the setting.
−
−
−
−
Setting is possible only up to a upper limit of the operating temperature range + 5 °C.
"
"
"
"
−
When the bath temperature arises above the
overtemperature switch-off point.
1. Double signal tone sounds
2. The display shows the indication for overtemperature
(TEMP) the fault triangle is flashing
!
heating is switched off on both poles,
!
pump and refrigeration system are switched off by the
electronics.
−
Rectify the cause of the fault.
−
Wait until the bath temperature has cooled down below the
switch-off point or set the switch-off point at a higher value.
When the display shows TEMP.
−
Reset with the key
☞
−
−
−
−
Before the unit is running unattended for longer periods
overtemperature protection
should be tested.
Therefore
−
Turn the potentiometer slowly anticlockwise.
!
The unit
must switch off at the bath temperature.
−
Step 1 - 2 (see above) must follow.
−
Set the overtemperature switch-off point again above the
bath temperature and wait until the indication TEMP
appears on the display, then
−
reset with the key.
Summary of Contents for Ecoline RE 204
Page 2: ......
Page 8: ...Brief operating instructions 2 ...
Page 59: ...Circuit diagram 10 08 99 YACE0060 53 11 Circuit diagram ...
Page 60: ...Circuit diagram 54 ...
Page 63: ...Pipe plan 16 08 99 YACE0060 57 12 Pipe plan ...
Page 64: ...Pipe plan 58 ...
Page 65: ...Pipe plan 16 08 99 YACE0060 59 ...
Page 68: ......