Safety notes
10.08.99
YACE0060
45
•
Depending on the bath liquid used and the mode of operation it is possible for toxic vapours to be
produced. Ensure appropriate ventilation!
•
When changing the bath liquid from water to oil, for temperatures above 100 °C, carefully remove all
traces of water, also from tubing and from the external consumer, otherwise
!
danger of burns
through delayed boiling!
•
Always pull out the mains plug before cleaning, maintenance or moving the thermostat!
•
Repairs on the control unit and the refrigeration system must be carried out by properly qualified
personnel only
•
Values for temperature control and indicating accuracy apply under normal conditions according to
DIN 58966. High-frequency electromagnetic fields may under special conditions lead to unfavourable
values. This does not affect the safety
Summary of Contents for Ecoline RE 204
Page 2: ......
Page 8: ...Brief operating instructions 2 ...
Page 59: ...Circuit diagram 10 08 99 YACE0060 53 11 Circuit diagram ...
Page 60: ...Circuit diagram 54 ...
Page 63: ...Pipe plan 16 08 99 YACE0060 57 12 Pipe plan ...
Page 64: ...Pipe plan 58 ...
Page 65: ...Pipe plan 16 08 99 YACE0060 59 ...
Page 68: ......