n
Ensure that adequate ventilation is provided
n
Minimum free room volume, room volume per kg of refrigerant
according to DIN 378-1
(For refrigerant quantity, see type plate or
and filling weight” on page 174).
Table 48: Refrigerant and room volume
Refrigerant
Room volume per kg of refrigerant
R-449A
2.81 m³/kg
R-452A
2.37 m³/kg
R-23
1.48 m³/kg
R-508B
4.0 m³/kg
Table 1
Unit
IN 130 T
IN 230 T
IN 230 TW
IN 530 T
ACC area
À
°C
-30 – 120
-30 – 120
-30 – 120
-30 – 120
Temperature stability
Á
K
±
0.05
±
0.05
±
0.05
±
0.05
Device dimensions
width x depth x height
mm
430 x 550 x 760 430 x 550 x 760 430 x 550 x 760 560 x 550 x
1325
Filling volume
‑
minimum
L
3.6
3.6
3.6
7.2
‑
maximum
L
8.7
8.7
8.7
20.5
Pump data 50 Hz
‑
maximum discharge pres-
sure
bar
3.5
3.5
3.5
3.5
‑
Maximum flow rate
l/min
40
40
40
40.0
Pump data 60 Hz
‑
maximum discharge pres-
sure
bar
4.0
4.0
4.0
4.6
‑
Maximum flow rate
l/min
45
45
45
45
Connector thread (outer)
inlet/outlet
Inches
G
¾
"
G
¾
"
G
¾
"
G
¾
"
Noise level 50 Hz
Â
dB(A)
61
63
58
62
Noise level 60 Hz
Â
dB(A)
61
63
60
66
Weight
kg
79.0
84.0
85.0
148.5
Distance between device and
environment
‑
Front
mm
500
500
200
500
‑
Back
mm
500
500
200
500
Installation location
Integral T
V6
Integral Process Thermostats and High-Temperature Thermostats
162 / 198