Table 1
Unit
IN 130 T
IN 230 T
IN 230 TW
IN 530 T
‑
Right
mm
500
500
200
500
‑
Left
mm
500
500
200
500
Table 2
Unit
IN 530 TW
IN 1030 T
IN 1330 TW
IN 1830 TW
ACC area
À
°C
-30 – 120
-30 – 150
-30 – 150
-30 – 150
Temperature stability
Á
K
±
0.05
±
0.1
±
0.1
±
0.1
Device dimensions
width x depth x height
mm
560 x 550 x
1325
760 x 650 x
1605
760 x 650 x
1605
760 x 650 x
1605
Filling volume
‑
minimum
L
7.2
9.7
9.7
9.7
‑
maximum
L
20.5
25.5
25.5
25.5
Pump data 50 Hz
‑
maximum discharge pres-
sure
bar
3.5
5.5
5.5
5.5
‑
Maximum flow rate
l/min
40.0
60.0
60.0
60.0
Pump data 60 Hz
‑
maximum discharge pres-
sure
bar
4.6
7.0
7.0
7.0
‑
Maximum flow rate
l/min
45
70.0
70.0
70.0
Connector thread (outer)
inlet/outlet
inch / mm G
¾
"
M38 x 1.5
M38 x 1.5
M38 x 1.5
Noise level 50 Hz
Â
dB(A)
62
69
59
64
Noise level 60 Hz
Â
dB(A)
64
70
62
67
Weight
kg
149.5
222.5
225.0
231.0
Distance between device and
environment
‑
Front
mm
200
500
200
200
‑
Back
mm
200
500
200
200
‑
Right
mm
200
500
200
200
‑
Left
mm
200
500
200
200
À
- ACC area (Active Cooling Control) according to DIN 12876 is the
working temperature range during operation with an active cooling unit.
Á
- Temperature stability determined according to standard DIN 12876-2
Â
- Noise level determined according to standard EN 11201 for operating
position in front of the device at 1 meter distance
V6
Integral Process Thermostats and High-Temperature Thermostats