Unit
IN 230 TW
IN 530 TW
IN 1330 TW
IN 1830 TW
Cooling water temperature
minimum/maximum
°C
10/30
10/30
10/30
10/30
Recommended cooling
water temperature
°C
15
15
15
15
Recommended cooling
water pressure difference
bar
3
3
3
3
Minimum cooling water
pressure difference*
bar
0.8
0.8
0.8
0.8
Maximum cooling water
pressure difference
bar
5
10
10
10
Maximum cooling water
pressure
bar
10
10
10
10
* only approved for a max. cooling water temperature of 20 °C
Table 51: Cooling output
Unit
IN 150 XT
IN 550 XT
IN 750 XT
Pump level
Cooling output at
200 °C
kW
1.50
5.00
7.00
8
100 °C
kW
1.50
5.00
7.00
8
20 °C
kW
1.50
5.00
7.00
8
10 °C
kW
1.50
4.80
7.00
8
0 °C
kW
1.30
4.60
5.40
8
-10 °C
kW
1.00
3.30
3.60
8
-20 °C
kW
0.70
2.30
2.60
4
-30 °C
kW
0.30
1.20
1.60
4
-40 °C
kW
0.06
0.50
0.80
4
-45 °C
kW
0.02
0.22
0.50
2
-50 °C
kW
---
0.10
---
2
Table 52: Cooling output
Unit
IN 250 XTW IN 550 XTW IN 950 XTW IN 1850
XTW
Pump level
Cooling output at (ambient
temperature 20 °C)
200 °C
kW
2.20
5.80
9.50
20.00
8
Integral XT with single-stage air-cooled
cooling unit
Integral XT with single-stage water-cooled
cooling unit
V6
Integral Process Thermostats and High-Temperature Thermostats
169 / 198