Personnel:
n
Specialized personnel
1.
Unscrew the covering plate on the right-hand side of the device.
2.
Unscrew the cover of the connection box.
3.
Undo the union nut of the cable gland.
4.
Insert the end of the mains cable through the cable gland.
5.
Pull the mains cable towards the terminals.
6.
Open the lever of the terminal as far as it will go.
7.
Insert the appropriate conductor in the terminal.
The terminals are suitable for solid conductors, fine-stranded
conductors and fine-stranded conductors with a ferrule.
8.
Close the lever until it first engages and continue until it finally
engages.
The conductor has been connected.
9.
Correctly connect the remaining conductors.
A ground terminal is mounted in the connection box in addi-
tion to the feed-through terminals. There is also a ground
stud in the connection box.
10. Tighten up the union nut of the cable gland.
11.
Screw the covers back onto the device.
Correct rotating field on the three-phase device
The device with a three-phase power supply checks that the rotating
field is correct when it is switched on and outputs a warning if
necessary.
n
The cable gland mounted on the device is suitable for cable diameters
22-31 mm.
A larger cable gland for cable diameters 26-35 mm is enclosed with the
device. Remove the bushing and replace the mounted cable gland with
the enclosed cable gland to install the thicker cable.
Installation of the mains cable in the device
Fig. 9: Covering plate in front of the con-
nection box
Fig. 10: Connection box cover
Fig. 11: Terminals in the connection box
Cable gland
V6
Integral Process Thermostats and High-Temperature Thermostats
26 / 198