Please note:
Water may not be used with any Integral XTs over the entire working
temperature range.
Table 6: Heat transfer liquid order numbers
Designation
Container size
Catalog number
5 L
10 L
20 L
Kryo 95
LZB 130
LZB 230
LZB 330
Kryo 71
LZB 131
LZB 231
LZB 331
Kryo 70
LZB 127
LZB 227
LZB 327
Kryo 65
LZB 118
LZB 218
LZB 318
Kryo 51
LZB 121
LZB 221
LZB 321
Kryo 30
LZB 109
LZB 209
LZB 309
Kryo 20
LZB 116
LZB 216
LZB 316
Aqua 90
LZB 120
LZB 220
LZB 320
Ultra 350
LZB 107
---
---
Ultra 301
LZB 153
LZB 253
LZB 353
4.5 Cooling water requirements
This section is relevant for:
n
Water-cooled devices
NOTICE!
Risk of cooling circuit leaking due to corrosion
Device damage
Do not use corrosive cooling water.
Cooling water is subject to specific purity requirements. A suitable procedure
must be employed to purify the cooling water in line with the contamination
in the water and maintain the water quality. Unsuitable cooling water may
cause the condenser and the entire cooling water circuit to become blocked
or damaged, or start to leak. The entire cooling circuit and cooling water
circuit may sustain extensive consequential damage as a result.